The process is to be done at the wbgeneral.
1) Fill a designated TMAH beaker with a TMAH solution in water; enough to submerge a wafer in a teflon TMAH single wafer holder or multiple wafers in a designated teflon boat.
2) Heat the solution to 95-100C on a wbgeneral hot plate.
3) While etching, cover the solution
with a designated glass cover.
NOTE: Most people use a 20% TMAH
solution. It etches at ~68 microns per hour, or through a wafer of standard
thickness (500 microns) in approximately 7-7.5 hours.
I. LONG
ETCHES:
The bath may be left over night.
II. PATTERNING:
Nitride and Oxide can be used
as a mask. Photoresist does not hold up to TMAH (it dissolves immediately).
III. STORAGE
AND REUSE:
You may save your solution in
a plastic bottle (cool it first) as long as it is properly labeled and in a
secondary containment (ie the bottom of the cart).
IV. DESIGNATED
TMAH BEAKERS AND HOLDERS:
If they need to be cleaned, use
a room temperture solution of 1:1:1 H20:H2O2:HCL
for 20 minutes and then rinse well. [See KOH cleanup
procedure]
The TMAH
quartzware is not to be used with metals or nonstandard materials because wafers
that get TMAH etched often go back into the main stream porcessing equiptment.
DO NOT USE GENERAL WETBENCH
beakers and holders etc... because if they are not designated for TMAH they
could have been used for anything -- including gold
|
_______________________________________________________________
Stanford Nanofabrication Facility Last Modified 09/03/2003 |