PROCESS ID: PUMP1125 DESCRIPTION: SAFE STANDBY UNDER VACUUM AT 250 DEG ST BANK: 3 TUBES: 11 REUSE: YES 0001.0000 IDLE 0001.0005 STEP UNLOAD=OFF 0001.0010 STEP LOAD=ON 0001.0015 STEP SPEED=15 0001.0020 STEP VACUUM=OFF 0001.0025 STEP N2PURG=1500 0001.0030 STEP TEMPL=250 0001.0035 STEP TEMPC=250 0001.0040 STEP TEMPS=250 0001.0045 STEP TCU LEARN=OFF 0005.0000 DOOR CHECK 0005.0005 STEP TIME:00:00:00 0005.0010 STEP IF DNTLK=ON GOTO 0010 0010.0000 PUMPDOWN 0010.0005 STEP TIME:00:04:00 0010.0010 STEP N2PURG=OFF 0010.0015 STEP N2PURG=0 0010.0020 STEP VACUUM=ON 0015.0000 GROSS LEAK CHECK 0015.0005 STEP TIME:00:00:10 0015.0010 STEP PRCPR TOLERANCE=1.0 0015.0015 STEP IF PRCPR>500 GOTO 0030 0020.0000 START PURGE FLOW 0020.0005 STEP TIME:00:01:00 0020.0010 STEP N2PURG=100 0020.0015 STEP TCU LEARN=OFF 0025.0000 STANDBY 0025.0005 STEP TIME:00:00:00 0025.0010 STEP IF PRCPR>500 GOTO 0030 0025.0015 STEP IF ALMACK=ON GOTO 0030 0030.0000 END 0030.0005 STEP END PROCESS