PROCESS ID: PPT620 DESCRIPTION: POLY FOR STRESS TESTING ST BANK: 3 TUBES: 9 REUSE: YES 0001.0000 IDLE 0001.0005 STEP VACUUM=OFF 0001.0010 STEP N2PURG=2000 0001.0015 STEP TEMPL=595 0001.0020 STEP TEMPC=600 0001.0025 STEP TEMPS=600 0001.0030 STEP UNLOAD=OFF 0001.0035 STEP LOAD=ON 0001.0040 STEP SPEED=25 0001.0045 STEP TCU LEARN=OFF 0005.0000 BOAT OUT 0005.0005 STEP TIME:00:00:00 0005.0010 STEP N2PURG=3500 0005.0015 STEP TEMPL=590 0005.0020 STEP TEMPC=600 0005.0025 STEP TEMPS=605 0005.0030 STEP LOAD=OFF 0005.0035 STEP UNLOAD=ON 0005.0040 STEP SPEED=25 0005.0045 STEP IF ALMACK=ON GOTO 0010 0005.0050 STEP TCU LEARN=OFF 0010.0000 BOAT IN 0010.0005 STEP TIME:00:00:00 0010.0010 STEP TEMPL=610 0010.0015 STEP TEMPC=610 0010.0020 STEP TEMPS=610 0010.0025 STEP UNLOAD=OFF 0010.0030 STEP LOAD=ON 0010.0035 STEP SPEED=25 0010.0040 STEP TCU CTLID=1 0010.0045 STEP TCU LEARN=OFF 0010.0050 STEP IF DNTLK=ON GOTO 0015 0015.0000 INITIAL PUMPDOWN 0015.0005 STEP TIME:00:03:00 0015.0010 STEP N2PURG=300 0015.0015 STEP VACUUM=ON 0015.0020 STEP TEMPL=617.0 0015.0025 STEP TEMPC=620.0 0015.0030 STEP TEMPS=623.0 0015.0035 STEP TCU CTLID=2 0015.0040 STEP TCU LEARN=OFF 0020.0000 START PURGE 0020.0005 STEP TIME:00:01:00 0020.0010 STEP N2PURG=150 0020.0015 STEP H2=60 0020.0020 STEP TCU CTLID=2 0020.0025 STEP TCU LEARN=OFF 0025.0000 STABILZE 0025.0005 STEP TIME:00:05:00 0025.0010 STEP TCU CTLID=2 0025.0015 STEP TCU LEARN=OFF 0025.0020 STEP IF PRCPR>1200 GOTO 0085 0030.0000 PUMP DOWN 0030.0005 STEP TIME:00:02:00 0030.0010 STEP N2PURG=OFF 0030.0015 STEP N2PURG=0 0030.0020 STEP H2=0 0030.0025 STEP H2=OFF 0030.0030 STEP TCU CTLID=2 0030.0035 STEP TCU LEARN=OFF 0035.0000 PUMPDOWN AND GROSS LEAK CHECK 0035.0005 STEP TIME:00:01:00 0035.0010 STEP TCU CTLID=2 0035.0015 STEP PRCPR TOLERANCE=1.0 0035.0020 STEP IF PRCPR>500 GOTO 0125 0040.0000 LEAK CHECK 0040.0005 STEP TIME:00:01:00 0040.0010 STEP TCU CTLID=2 0040.0015 STEP VACUUM=OFF 0040.0020 STEP TCU LEARN=OFF 0040.0025 STEP IF PRCPR>60 GOTO 0125 0045.0000 REPUMP 0045.0005 STEP TIME:00:02:00 0045.0010 STEP VACUUM=ON 0045.0015 STEP N2PURG=200 0045.0020 STEP TCU CTLID=2 0045.0025 STEP TCU LEARN=OFF 0050.0000 START FLOW 0050.0005 STEP TIME:00:01:00 0050.0010 STEP N2PURG=135 0050.0015 STEP H2=80 0050.0020 STEP TCU CTLID=2 0050.0025 STEP TCU LEARN=OFF 0055.0000 PRCPR STABLIZE 0055.0005 STEP TIME:00:02:00 0055.0010 STEP PRCPR=400 0055.0015 STEP TCU CTLID=2 0055.0020 STEP TCU LEARN=OFF 0060.0000 FINAL STABILIZE TO +-1 DEG 0060.0005 STEP TIME:00:30:00 0060.0010 STEP H2=110 0060.0015 STEP TCU CTLID=2 0060.0020 STEP TCU LEARN=OFF 0060.0025 STEP IF PRCPR>1200 GOTO 0085 0065.0000 *DEPOSITION* 0065.0005 STEP TIME:VARIABLE 0065.0010 STEP N2PURG=OFF 0065.0015 STEP N2PURG=0 0065.0020 STEP SIH4=136 0065.0025 STEP H2=110 0065.0030 STEP TCU CTLID=2 0065.0035 STEP TCU LEARN=OFF 0065.0040 STEP IF PRCPR>1200 GOTO 0085 0070.0000 H2 PURGE 0070.0005 STEP TIME:00:00:30 0070.0010 STEP SIH4=0 0070.0015 STEP SIH4=OFF 0070.0020 STEP TCU CTLID=3 0070.0025 STEP TCU LEARN=OFF 0070.0030 STEP IF PRCPR>1200 GOTO 0085 0075.0000 PUMP DOWN 0075.0005 STEP TIME:00:02:00 0075.0010 STEP H2=OFF 0075.0015 STEP H2=0 0075.0020 STEP PRCPR=OFF 0075.0025 STEP PRCPR=0 0075.0030 STEP TCU LEARN=OFF 0080.0000 PURGE 0080.0005 STEP TIME:00:05:00 0080.0010 STEP N2PURG=150 0080.0015 STEP TCU LEARN=OFF 0085.0000 1ST PUMP DOWN 0085.0005 STEP TIME:00:02:00 0085.0010 STEP N2PURG=OFF 0085.0015 STEP N2PURG=0 0085.0020 STEP TEMPL=500 0085.0025 STEP TEMPC=500 0085.0030 STEP TEMPS=500 0085.0035 STEP TCU LEARN=OFF 0090.0000 1ST PURGE 0090.0005 STEP TIME:00:01:00 0090.0010 STEP N2PURG=300 0090.0015 STEP TCU LEARN=OFF 0095.0000 2ND PUMPDOWN 0095.0005 STEP TIME:00:01:00 0095.0010 STEP N2PURG=OFF 0095.0015 STEP N2PURG=0 0095.0020 STEP TCU LEARN=OFF 0100.0000 2ND PURGE 0100.0005 STEP TIME:00:01:00 0100.0010 STEP N2PURG=350 0100.0015 STEP TCU LEARN=OFF 0105.0000 PUMP DOWN 0105.0005 STEP TIME:00:01:00 0105.0010 STEP N2PURG=OFF 0105.0015 STEP N2PURG=0 0105.0020 STEP TCU LEARN=OFF 0110.0000 3RD PURGE 0110.0005 STEP TIME:00:01:00 0110.0010 STEP N2PURG=350 0110.0015 STEP TCU LEARN=OFF 0115.0000 FINAL PUMP DOWN 0115.0005 STEP TIME:00:01:00 0115.0010 STEP N2PURG=OFF 0115.0015 STEP N2PURG=0 0115.0020 STEP TCU LEARN=OFF 0120.0000 FINAL PURGE AND RAMPDOWN 0120.0005 STEP TIME:00:00:00 0120.0010 STEP N2PURG=400 0120.0015 STEP TEMPL TOLERANCE=1.0 0120.0020 STEP IF TEMPL<600 GOTO 0125 0120.0025 STEP TCU LEARN=OFF 0125.0000 SLOW BACKFILL AND ABORT STEP 0125.0005 STEP TIME:00:08:00 0125.0010 STEP VACUUM=OFF 0125.0015 STEP TEMPL=550 0125.0020 STEP TEMPC=550 0125.0025 STEP TEMPS=550 0125.0030 STEP N2PURG=2250 0125.0035 STEP N2BKFL=750 0125.0040 STEP TCU LEARN=OFF 0130.0000 BOAT OUT 0130.0005 STEP TIME:00:00:00 0130.0010 STEP N2BKFL=OFF 0130.0015 STEP N2BKFL=0 0130.0020 STEP N2PURG=3500 0130.0025 STEP TEMPL=500 0130.0030 STEP TEMPC=525 0130.0035 STEP TEMPS=550 0130.0040 STEP LOAD=OFF 0130.0045 STEP UNLOAD=ON 0130.0050 STEP SPEED=25 0130.0055 STEP IF ALMACK=ON GOTO 0135 0130.0060 STEP TCU LEARN=OFF 0135.0000 END 0135.0005 STEP END PROCESS