PROCESS ID: LSIN950A DESCRIPTION: ATTEMPT TO FORM BORON DOPED NITRIDE ST BANK: 2 TUBES: 5 REUSE: YES 0001.0000 IDLE 0001.0005 STEP TEMPL=800 0001.0010 STEP TEMPC=800 0001.0015 STEP TEMPS=800 0001.0020 STEP UNLOAD=OFF 0001.0025 STEP LOAD=ON 0001.0030 STEP SPEED=10 0001.0035 STEP N2=2.5 0001.0040 STEP N2CARR=150 0005.0000 DOOR OPEN 0005.0005 STEP TIME:00:00:00 0005.0010 STEP N2=5.0 0005.0015 STEP TEMPL=725 0005.0020 STEP TEMPC=775 0005.0025 STEP TEMPS=800 0005.0030 STEP TCU LEARN=OFF 0005.0035 STEP LOAD=OFF 0005.0040 STEP UNLOAD=ON 0005.0045 STEP SPEED=20 0005.0050 STEP IF ALMACK=ON GOTO 0010 0010.0000 BOAT IN AND STABILIZE TO +-5 DEG 0010.0005 STEP TIME:00:20:00 0010.0010 STEP TEMPL=800 0010.0015 STEP TEMPC=800 0010.0020 STEP TEMPS=800 0010.0025 STEP UNLOAD=OFF 0010.0030 STEP LOAD=ON 0010.0035 STEP SPEED=12 0010.0040 STEP TCU CTLID=1 0010.0045 STEP TCU LEARN=OFF 0015.0000 RAMP UP AND FINAL STABILIZE TO +-1.0 DEG 0015.0005 STEP TIME:00:25:00 0015.0010 STEP TEMPL=950 0015.0015 STEP TEMPC=950 0015.0020 STEP TEMPS=950 0015.0025 STEP N2=8.2 0015.0030 STEP O2LO=50 0015.0035 STEP TCU CTLID=12 0015.0040 STEP TCU LEARN=OFF 0020.0000 BEGIN CARRIER FLOW 0020.0005 STEP TIME:00:00:30 0020.0010 STEP N2CARR=125 0020.0015 STEP N2=8.1 0020.0020 STEP O2LO=51 0020.0025 STEP TCU CTLID=12 0020.0030 STEP TCU LEARN=OFF 0025.0000 BEGIN SOURCE FLOW 0025.0005 STEP TIME:00:00:30 0025.0010 STEP BBR3=ON 0025.0015 STEP TCU CTLID=12 0025.0020 STEP TCU LEARN=OFF 0030.0000 DOPE 0030.0005 STEP TIME:00:30:00 0030.0010 STEP TUBE=ON 0030.0015 STEP TCU CTLID=12 0030.0020 STEP TCU LEARN=OFF 0035.0000 PURGE 0035.0005 STEP TIME:00:05:00 0035.0010 STEP BBR3=OFF 0035.0015 STEP TCU CTLID=12 0035.0020 STEP TCU LEARN=OFF 0040.0000 RAMPDOWN 0040.0005 STEP TIME:00:25:00 0040.0010 STEP TUBE=OFF 0040.0015 STEP O2LO=200 0040.0020 STEP N2=4.5 0040.0025 STEP TEMPL=750 0040.0030 STEP TEMPC=750 0040.0035 STEP TEMPS=750 0040.0040 STEP TCU LEARN=OFF 0045.0000 PURGE 0045.0005 STEP TIME:00:01:00 0045.0010 STEP O2LO=OFF 0045.0015 STEP O2LO=0 0045.0020 STEP TCU LEARN=OFF 0050.0000 BOAT OUT 0050.0005 STEP TIME:00:00:00 0050.0010 STEP TEMPL=700 0050.0015 STEP TEMPC=750 0050.0020 STEP TEMPS=775 0050.0025 STEP TCU LEARN=OFF 0050.0030 STEP LOAD=OFF 0050.0035 STEP UNLOAD=ON 0050.0040 STEP SPEED=15 0050.0045 STEP IF ALMACK=ON GOTO 0055 0055.0000 END 0055.0005 STEP END PROCESS