PROCESS ID: CMZW1150 DESCRIPTION: STEAM AT 1150 RAMP WITH O2 ONLY BANK: 1 TUBES: 7 REUSE: YES 0001.0000 IDLE 0001.0005 STEP N2=2.5 0001.0010 STEP O2HI=OFF 0001.0015 STEP O2HI=0 0001.0020 STEP TEMPL=800 0001.0025 STEP TEMPC=800 0001.0030 STEP TEMPS=800 0001.0035 STEP UNLOAD=OFF 0001.0040 STEP LOAD=ON 0001.0045 STEP SPEED=10 0005.0000 BOAT PULL 0005.0005 STEP TIME:00:00:00 0005.0010 STEP N2=5.0 0005.0015 STEP TEMPL=700 0005.0020 STEP TEMPC=750 0005.0025 STEP TEMPS=800 0005.0030 STEP LOAD=OFF 0005.0035 STEP UNLOAD=ON 0005.0040 STEP SPEED=8 0005.0045 STEP IF ALMACK=ON GOTO 0010 0010.0000 BOAT IN 0010.0005 STEP TIME:00:20:00 0010.0010 STEP TEMPL=800 0010.0015 STEP TEMPC=800 0010.0020 STEP TEMPS=800 0010.0025 STEP UNLOAD=OFF 0010.0030 STEP LOAD=ON 0010.0035 STEP SPEED=8 0010.0040 STEP TCU CTLID=1 0010.0045 STEP TCU LEARN=ON 0015.0000 RAMP UP TO 1150 DEG +- .5 0015.0005 STEP TIME:00:50:00 0015.0010 STEP TEMPL=1150 0015.0015 STEP TEMPC=1150 0015.0020 STEP TEMPS=1150 0015.0025 STEP N2=OFF 0015.0030 STEP N2=0 0015.0035 STEP O2HI=ON 0015.0040 STEP O2HI=2500 0015.0045 STEP TCU CTLID=8 0015.0050 STEP TCU LEARN=OFF 0020.0000 DRY OX 0020.0005 STEP TIME:00:05:00 0020.0010 STEP N2=OFF 0020.0015 STEP N2=0 0020.0020 STEP O2LO=OFF 0020.0025 STEP O2LO=0 0020.0030 STEP O2HI=2500 0020.0035 STEP TCU CTLID=8 0020.0040 STEP TCU LEARN=OFF 0025.0000 STEAM 0025.0005 STEP TIME:VARIABLE 0025.0010 STEP H2/O2=150 0025.0015 STEP O2HI=1000 0025.0020 STEP TCU CTLID=27 0025.0025 STEP TCU LEARN=OFF 0030.0000 DRY OX 0030.0005 STEP TIME:00:05:00 0030.0010 STEP H2/O2=OFF 0030.0015 STEP H2/O2=0 0030.0020 STEP TCU CTLID=8 0030.0025 STEP TCU LEARN=OFF 0035.0000 ANNEAL 0035.0005 STEP TIME:VARIABLE 0035.0010 STEP O2HI=OFF 0035.0015 STEP O2HI=0 0035.0020 STEP N2=ON 0035.0025 STEP N2=2.4 0035.0030 STEP TCU CTLID=8 0035.0035 STEP TCU LEARN=OFF 0040.0000 RAMP DOWN 0040.0005 STEP TIME:00:00:00 0040.0010 STEP O2HI=OFF 0040.0015 STEP O2HI=0 0040.0020 STEP N2=5.0 0040.0025 STEP TEMPL=750 0040.0030 STEP TEMPC=750 0040.0035 STEP TEMPS=750 0040.0040 STEP IF ALMACK=ON GOTO 0045 0045.0000 BOAT OUT 0045.0005 STEP TIME:00:00:00 0045.0010 STEP TEMPL=700 0045.0015 STEP TEMPC=750 0045.0020 STEP TEMPS=775 0045.0025 STEP TCU LEARN=OFF 0045.0030 STEP LOAD=OFF 0045.0035 STEP UNLOAD=ON 0045.0040 STEP SPEED=8 0045.0045 STEP IF ALMACK=ON GOTO 0050 0050.0000 END 0050.0005 STEP END PROCESS