GO DIRECT TO SPECIFIC BENCHES:

Diffusion Wet Bench (wbdiff)
General Use Wet Bench (wbgeneral)
Silicide Wet Bench (wbsilicide)


STANDARD PREDIFFUSION CLEAN INFORMATION

WHO NEEDS IT?

  1. Wafers should not have metal on them when getting this clean. Depending on where the clean will be done and where the wafers are going after, the wafers could have had the metal removed or had it turned to silicide prior to this clean (see section below "Where should I do it?" for details).
  2. Standard prediffusion clean should be done on wafers that are going into furnaces, LPCVD systems or metal deposition.
  3. Wafers should be loaded into the furnace or deposition within one hour of the clean process. If they sit longer than one hour they MUST be RECLEANED.
  4. Wafers with resist on them should get resist strip before going into prediffusion clean.
  5. IMPORTANT NOTE: Never touch your wafers. If you do they must be thrown away because there is no clean that can remove the contaminants left by fingerprints.

WHAT IS THE PROCESS

VARIATIONS?

  1. If you are going to metal deposition or an LPCVD deposition you should do the HF dip last. This prevents an oxide from being between the silicon and the metal deposition. It prevents the LPCVD films from having a HAZY deposition.
  2. In the previous case, some people prefer to do two HF dips (the one that is in the middle of the clean process to remove the chemical oxide before going into the 5:1:1 H2O:H2O2:HCl bath AND to do the HF dip last to get the chemical oxide off before metal deposition and LPCVD.
  3. When cleaning thin very thin gate oxides, you may want to minimize or eliminate the HF dip step (to preserve the oxide), but keep in mind the possibility of bad contact to metal layers or hazy depositions.

HOW DOES IT WORK?

  1. The H2SO4:H2O2 (Sulfuric Peroxide) solution removes the organic contaminants from the wafers.
  2. The HF dip removes the chemical oxide put on during the Sulfuric Peroxide step.
  3. The H2O:HCl:H2O2 removes any metal contaminants that might be on the wafer (it is most effective if done after stripping off the chemical oxide).
  4. When an HF dip is done last it removes the chemical oxide put on by the previous clean step.

WHERE SHOULD I DO IT?

  1. The process is used to clean wafers that do not have metal on them.
  2. If the wafers have never had metal on them or been in any metal contaminated equipment the clean would be done at the wbdiff (,4", 3", 2" wafers) OR at wbgeneral in designated nonmetal quartz beakers (nonstandard sized wafers or pieces of wafers). Wafers cleaned this way could go into nonmetal furnaces (i.e. tylan1-6, tylanbpsg (nonmetal side), tylanpoly, tylannitride).
  3. If they had "standard metals" on them but it was stripped off, the wafers would get cleaned at the wbsilicide. (4", 3", 2" wafers) OR wbgeneral in designated std metal ONLY quartz beakers (nonstandard sized wafers or pieces of wafers). NOTE: "standard metals" are Al, W and Ti. Wafers cleaned this way could go into "standard metal" furnaces ONLY (i.e. tylan7, tylanfga, tylanbpsg (metal side only).
  4. NOTE: There is a cassette available at wgdiff, wbsilicide, wbnonmetal and wbmetal for cleaning 6 " wafers.
  5. NON-Standard Metal: If your wafers had non-standard metals (Al, Ti and W are "standard metals" so all other metals are non-standard) on them or were exposed to equipment or beakers that are contaminated with non-standard metals you should clean them at the wbgeneral in your own clearly labeled equipment that allows those materials (see Materials allowed for details). NOTE: the gold designated Pyrex beakers at the wbgeneral may be used if the equipment you need to use allows gold and other non-standard metals (do not do an HF dip in Pyrex beakers though because you will get contaminants from the beakers on your wafers -- sodium, boron, etc....).