01: Problem tylan1: reset connector ((none))
02: Problem tylan1: Tylan1 MFC alarm sounded during TCACLEAN, Alarm acknowledge silenced it, alarm did not recur ((none))
03: Problem tylan1: reset mfs 460 ((none))
04: Problem tylan1: led display is out ((none))
05: Problem tylan1: Did not get up to temp?? (ericp.snf.stanford.edu)
06: Problem Clear tylan1: Reset TCU (gaul.snf.stanford.edu)
07: Problem Clear tylan1: O.K. now (gaul.snf.stanford.edu)
08: Problem tylan1: Fur status Indicator off (price.snf.stanford.edu)
09: Problem Clear tylan1: looks good! (latta.snf.stanford.edu)
10: Shutdown tylan1: (kalavade.snf.stanford.edu)
11: Shutdown Clear tylan1: Reset breaker (gaul.snf.stanford.edu)
12: Problem tylan1: (tui3.snf.stanford.edu)
13: Shutdown tylan1: tube stuck! (emc.snf.stanford.edu)
14: Shutdown Clear tylan1: Replaced bushings (gaul.snf.stanford.edu)
15: Problem Clear tylan1: Replaced bushings (gaul.snf.stanford.edu)
16: Shutdown tylan1: Door interlock not being made (gaul.snf.stanford.edu)
17: Problem tylan1: TCU error (venkat.snf.stanford.edu)
18: Problem tylan1: reset the temp controller (seymour.snf.stanford.edu)
19: Problem tylan1 2001-01-10 11:09:38: mfs alarm running constantly in ready mode (lindaw.snf.stanford.edu)
20: Problem tylan1 2001-01-19 18:54:58: No O2 gas flow in the chamber? (pzwahlen.snf.stanford.edu)
21: Re: Problem tylan1 2001-01-10 11:09:38: mfs alarm running constantly in ready mode (seymour.snf.stanford.edu)
22: Re: Problem tylan1 2001-01-19 18:54:58: O2 flow (seymour.snf.stanford.edu)
23: Problem tylan1 2001-03-19 10:01:42: Display is dim and has bad connection (king.snf.stanford.edu)
24: Re: Problem tylan1 2001-03-19 10:01:42: Display is dim and has bad connection (seymour.snf.stanford.edu)
25: Shutdown tylan1 2001-06-10 18:12:48: no O2 (aaronp.snf.stanford.edu)
26: Re: Shutdown tylan1 2001-06-10 18:12:48: no O2 (seymour.snf.stanford.edu)
27: Shutdown tylan1 2001-07-12 16:47:43: center zone is out. (caudillo.snf.stanford.edu)
28: Re: Shutdown tylan1 2001-07-12 16:47:43: center zone is out. (seymour.snf.stanford.edu)
29: Problem tylan1 2001-07-14 18:17:04: TEMPL runs low (ericp.snf.stanford.edu)
30: Shutdown tylan1 2001-07-16 13:17:12: load zone power problem (seymour.snf.stanford.edu)
31: Re: Shutdown tylan1 2001-07-16 13:17:12: load zone power problem (seymour.snf.stanford.edu)
32: Problem tylan1 2001-07-26 18:35:00: tylan1:15% off the target thickness, front is 20% off (gladys.snf.stanford.edu)
33: Shutdown tylan1 2001-08-01 09:21:28: need to address temperature problem (seymour.snf.stanford.edu)
34: Re: Shutdown tylan1 2001-08-01 09:21:28: need to address temperature problem (seymour.snf.stanford.edu)
35: Re: Problem tylan1 2001-07-26 18:35:00: tylan1:15% off the target thickness, front is 20% off (seymour.snf.stanford.edu)
36: Re: Problem tylan1 2001-07-14 18:17:04: TEMPL runs low (seymour.snf.stanford.edu)
37: Shutdown tylan1 2001-08-25 07:37:09: (seymour.snf.stanford.edu)
38: Re: Shutdown tylan1 2001-08-25 07:37:09: (seymour.snf.stanford.edu)
39: Problem tylan1 2001-09-26 10:57:54: tylan1: showing an OT error (gladys.snf.stanford.edu)
40: Shutdown tylan1 2001-09-26 11:58:36: tube overtemped (seymour.snf.stanford.edu)
41: Re: Shutdown tylan1 2001-09-26 11:58:36: tube overtemped (seymour.snf.stanford.edu)
42: Re: Problem tylan1 2001-09-26 10:57:54: tylan1: showing an OT error (seymour.snf.stanford.edu)
43: Shutdown tylan1 2001-09-27 08:02:36: waiting for testrun (gladys.snf.stanford.edu)
44: Re: Shutdown tylan1 2001-09-27 08:02:36: waiting for testrun (seymour.snf.stanford.edu)
45: Shutdown tylan1 2001-10-23 15:07:55: needs tca clean and thickness testing (seymour.snf.stanford.edu)
46: Re: Shutdown tylan1 2001-10-23 15:07:55: needs tca clean and thickness testing (latta.snf.stanford.edu)
47: Shutdown tylan1 2001-11-29 08:02:11: replacing the rear baffle (seymour.snf.stanford.edu)
48: Re: Shutdown tylan1 2001-11-29 08:02:11: replacing the rear baffle (seymour.snf.stanford.edu)
49: Shutdown tylan1 2001-12-03 15:36:21: Tube doesn't pull out + bad TC (amol.snf.stanford.edu)
50: Re: Shutdown tylan1 2001-12-03 15:36:21: Tube doesn't pull out + bad TC (seymour.snf.stanford.edu)
51: Problem tylan1 2002-01-28 21:55:20: H2 MFC alarm (ericp.snf.stanford.edu)
52: Re: Problem tylan1 2002-01-28 21:55:20: H2 MFC alarm (seymour.snf.stanford.edu)
53: Shutdown tylan1 2002-02-05 17:09:41: Tylan Bank 1 move (gladys.snf.stanford.edu)
54: Shutdown tylan1 2002-02-28 20:47:47: load motor shorted in out position (ericp.snf.stanford.edu)
55: Re: Shutdown tylan1 2002-02-28 20:47:47: load motor shorted in out position (latta.snf.stanford.edu)
56: Shutdown tylan1 2002-06-05 11:30:32: element wire connection bad (seymour.snf.stanford.edu)
57: Re: Shutdown tylan1 2002-06-05 11:30:32: element wire connection bad (seymour.snf.stanford.edu)
58: Problem tylan1 2002-07-05 11:33:47: TCACLEAN due to possible glass contamination (ychliu.snf.stanford.edu)
59: Shutdown tylan1 2002-07-08 10:15:42: sodium contaminated (gladys.snf.stanford.edu)
60: Re: Shutdown tylan1 2002-07-08 10:15:42: sodium contaminated (seymour.snf.stanford.edu)
61: Re: Problem tylan1 2002-07-05 11:33:47: TCACLEAN due to possible glass contamination (seymour.snf.stanford.edu)
62: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation (rajesh.snf.stanford.edu)
63: Re: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation (John Shott)
64: Re: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation (Nancy Latta)
65: Shutdown tylan1 2002-08-28 12:04:30: tube needs to be cleaned (seymour.snf.stanford.edu)
66: Re: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation (Jim McVittie)
67: Re: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation (Dick Crane)
68: Use of Steam in Ox Furnaces (Mary Tang)
69: Re: Shutdown tylan1 2002-08-28 12:04:30: tube needs to be cleaned---Done (mcvittie.snf.stanford.edu)
70: Re: Problem tylan1 2002-08-28 10:52:08: Can't do oxidation --- Parts replaced (mcvittie.snf.stanford.edu)
71: Problem tylan1 2002-08-29 19:32:50: Waiting for changes in TCA system (mcvittie.snf.stanford.edu)
72: Problem tylan1 2002-09-01 14:56:35: See Ty1 Equip email achieve note (mcvittie.snf.stanford.edu)
73: Shutdown tylan1 2002-09-06 08:33:19: doing a quick check (seymour.snf.stanford.edu)
74: Re: Shutdown tylan1 2002-09-06 08:33:19: doing a quick check (seymour.snf.stanford.edu)
75: Shutdown tylan1 2002-11-05 14:23:12: cooling down for TCA mod and pull and clean (seymour.snf.stanford.edu)
76: Test run before pull and clean (Nancy Latta)
77: Re: Shutdown tylan1 2002-11-05 14:23:12: cooling down for TCA mod and pull and clean (seymour.snf.stanford.edu)
78: Re: Problem tylan1 2002-08-29 19:32:50: Waiting for changes in TCA system (seymour.snf.stanford.edu)
79: Shutdown tylan1 2002-11-08 06:45:17: waiting for test on uniformity and thickness (seymour.snf.stanford.edu)
80: Comment tylan1 2002-11-08 10:16:59: There are no comments on the test yesterday (seymour.snf.stanford.edu)
81: 11/7/results wet ox look good (Maurice Stevens)
82: Re: Shutdown tylan1 2002-11-08 06:45:17: waiting for test on uniformity and thickness (maurice.snf.stanford.edu)
83: Re: Problem tylan1 2002-09-01 14:56:35: See Ty1 Equip email achieve note (seymour.snf.stanford.edu)
84: Problem tylan1 2002-11-14 15:23:36: Torch baffle cracked (latta.snf.stanford.edu)
85: Re: Problem tylan1 2002-11-14 15:23:36: Torch baffle cracked (latta.snf.stanford.edu)
86: Problem tylan1 2002-12-15 21:33:46: bad uniformity (right boat, 3200--4400A) (yyao.snf.stanford.edu)
87: Shutdown tylan1 2002-12-18 09:48:58: Bad uniformity (tberg.snf.stanford.edu)
88: Uniformity test after bad uniformity reported (Nancy Latta)
89: Re: Shutdown tylan1 2002-12-18 09:48:58: Bad uniformity: testrun looks good (gladys.snf.stanford.edu)
90: Re: Problem tylan1 2002-12-15 21:33:46: bad uniformity (right boat, 3200--4400A)reported twice (gladys.snf.stanford.edu)
91: Shutdown tylan1 2003-01-08 23:13:00: tube doesn't come out / quartz pieces in tube (hector.snf.stanford.edu)
92: Re: Shutdown tylan1 2003-01-08 23:13:00: tube doesn't come out / quartz pieces in tube (mdickey.snf.stanford.edu)
93: Problem tylan1 2003-01-09 11:41:28: Sticks during tube pull (bchui.snf.stanford.edu)
94: Re: Problem tylan1 2003-01-09 11:41:28: Sticks during tube pull (mdickey.snf.stanford.edu)
95: Problem tylan1 2003-01-11 13:26:04: broken wafer flakes on boats & rods (hector.snf.stanford.edu)
96: Shutdown tylan1 2003-01-13 06:51:14: Orings flaking (tberg.snf.stanford.edu)
97: Problem tylan1 2003-01-13 11:30:18: O ring replaced (mdickey.snf.stanford.edu)
98: Re: Shutdown tylan1 2003-01-13 06:51:14: Orings flaking:tca'd and tested..ok to run (gladys.snf.stanford.edu)
99: Re: Problem tylan1 2003-01-13 11:30:18: O ring replaced...tca'd (gladys.snf.stanford.edu)