Messages 0xx


01: Problem tylan4: carr mfc not off at step 15 ((none))
02: Problem Clear tylan4: reset zero on mfc ((none))
03: Problem tylan4: tylan 4 possible organic contamination (greene.snf.stanford.edu)
04: Problem Clear tylan4: Turns out not contaminated (gaul.snf.stanford.edu)
05: Shutdown tylan4: quartz sheath on cantilever shattered (sminne.snf.stanford.edu)
06: Shutdown Clear tylan4: Cantilever sheaths replaced (gaul.snf.stanford.edu)
07: Problem tylan4: steam didn't turn on (Bill Price) (grupp.snf.stanford.edu)
08: Shutdown tylan4: Cooling fan problem (gaul.snf.stanford.edu)
09: Problem Clear tylan4: shutdown to fix (gaul.snf.stanford.edu)
10: Shutdown Clear tylan4: cooling fan replaced (gaul.snf.stanford.edu)
11: Problem tylan4: door will not open (sminne.snf.stanford.edu)
12: Problem tylan4: crack in rear quartz support arm (sminne.snf.stanford.edu)
13: Shutdown tylan4: Rear quartz sheath cracked. Shutdown per keith. (sminne.snf.stanford.edu)
14: Shutdown Clear tylan4: completed (gaul.snf.stanford.edu)
15: Problem Clear tylan4: completed (gaul.snf.stanford.edu)
16: Problem Clear tylan4: completed (gaul.snf.stanford.edu)
17: Problem tylan4: Did not unload after stp 40 of wet1100 (caudillo.snf.stanford.edu)
18: Problem tylan4: Tube does not go in all the way causing a door interlock failure. Can manually push tube all the way in. (bowers.snf.stanford.edu)
19: Shutdown tylan4: Fault on door intrerlock. Tube does not go all the way in. Must manually push tube in. (bowers.snf.stanford.edu)
20: Problem Clear tylan4: loader motor replaced (seymour.snf.stanford.edu)
21: Shutdown Clear tylan4: replaced loader motor (seymour.snf.stanford.edu)
22: Problem Clear tylan4: replaced loader motor (seymour.snf.stanford.edu)
23: Problem tylan4: Must manually push tube in. (bowers.snf.stanford.edu)
24: Problem Clear tylan4: (seymour.snf.stanford.edu)
25: Shutdown tylan4: Tylan 4 (kenney.snf.stanford.edu)
26: Shutdown Clear tylan4: (seymour.snf.stanford.edu)
27: Problem tylan4: boat sticks during boat in (greene.snf.stanford.edu)
28: Comment tylan4: loader problems (seymour.snf.stanford.edu)
29: Problem Clear tylan4: loader problems (seymour.snf.stanford.edu)
30: Shutdown tylan4: push/pull sticks when opening/closing (dhelqaq.snf.stanford.edu)
31: Shutdown Clear tylan4: sticking loader (seymour.snf.stanford.edu)
32: Problem tylan4: tube had trouble going into furnace (skoh.snf.stanford.edu)
33: Problem tylan4: Push-Puller sticky (price.snf.stanford.edu)
34: Problem tylan4: Tube will not close (price.snf.stanford.edu)
35: Problem Clear tylan4: replaced loader bearings (seymour.snf.stanford.edu)
36: Problem Clear tylan4: (seymour.snf.stanford.edu)
37: Problem Clear tylan4: (seymour.snf.stanford.edu)
38: Shutdown tylan4: baffle is broken (lindaw.snf.stanford.edu)
39: Shutdown Clear tylan4: replaced baffel (seymour.snf.stanford.edu)
40: Problem tylan4: (davehong.snf.stanford.edu)
41: Problem tylan4: (davehong.snf.stanford.edu)
42: Problem tylan4: (seymour.snf.stanford.edu)
43: Problem tylan4: (seymour.snf.stanford.edu)
44: Comment tylan4: adjusted zero for N2 carrier mfc ok now (seymour.snf.stanford.edu)
45: Shutdown tylan4: carrier N2 and lo O2 reading very high with (seymour.snf.stanford.edu)
46: Re: Shutdown tylan4 2000-12-10 22:27:36: The MFC controller has been installed (seymour.snf.stanford.edu)
47: Shutdown tylan4 2001-02-28 11:40:46: SCR package overheating, tube down (seymour.snf.stanford.edu)
48: Re: Shutdown tylan4 2001-02-28 11:40:46: SCR fan replaced (seymour.snf.stanford.edu)
49: Problem tylan4 2001-03-03 11:03:30: MFS alarming and alarming, and alarming...... (lindaw.snf.stanford.edu)
50: Problem tylan4 2001-03-07 20:13:24: tylan 4 found in manual mode. (caudillo.snf.stanford.edu)
51: Re: Problem tylan4 2001-03-07 20:13:24: tylan 4 found in manual mode. (seymour.snf.stanford.edu)
52: Shutdown tylan4 2001-03-22 09:59:45: mfc's (mdickey.snf.stanford.edu)
53: Comment tylan4 2001-03-22 11:05:51: Sensor (mdickey.snf.stanford.edu)
54: Re: Shutdown tylan4 2001-03-22 09:59:45: mfc's (mdickey.snf.stanford.edu)
55: Re: Problem tylan4 2001-03-03 11:03:30: MFS alarming and alarming, and alarming...... (mdickey.snf.stanford.edu)
56: Shutdown tylan4 2001-03-22 13:22:01: H2 flow problem (seymour.snf.stanford.edu)
57: Re: Shutdown tylan4 2001-03-22 13:22:01: H2 flow problem (seymour.snf.stanford.edu)
58: Shutdown tylan4 2002-01-22 13:32:05: needs testing after repair (seymour.snf.stanford.edu)
59: Re: Shutdown tylan4 2002-01-22 13:32:05: needs testing after repair (seymour.snf.stanford.edu)
60: Shutdown tylan4 2002-02-05 17:10:32: Tylan Bank 1 move (gladys.snf.stanford.edu)
61: Shutdown tylan4 2002-03-07 15:12:24: Boat doesn't go in all the way. (latta.snf.stanford.edu)
62: Re: Shutdown tylan4 2002-02-05 17:10:32: Tylan Bank 1 move (seymour.snf.stanford.edu)
63: Re: Shutdown tylan4 2002-03-07 15:12:24: Boat doesn't go in all the way. (shott.snf.stanford.edu)
64: Shutdown tylan4 2002-03-11 09:25:50: needs test (seymour.snf.stanford.edu)
65: Re: Shutdown tylan4 2002-03-11 09:25:50: needs test (latta.snf.stanford.edu)
66: Problem tylan4 2002-04-15 16:49:35: black "o-ring" on door seems to have melted (beckwith.snf.stanford.edu)
67: Shutdown tylan4 2002-04-16 15:51:56: Bad ox uniformity after O-ring was replaced. (gladys.snf.stanford.edu)
68: Re: Shutdown tylan4 2002-04-16 15:51:56: Bad ox uniformity after O-ring was replaced. (gladys.snf.stanford.edu)
69: Re: Problem tylan4 2002-04-15 16:49:35: black "o-ring" on door seems to have melted (gladys.snf.stanford.edu)
70: Shutdown tylan4 2002-05-23 15:23:30: H2 Flow problem (seymour.snf.stanford.edu)
71: Re: Shutdown tylan4 2002-05-23 15:23:30: H2 Flow problem (seymour.snf.stanford.edu)
72: Shutdown tylan4 2002-05-24 09:59:58: needs test after TCA Clean finished (seymour.snf.stanford.edu)
73: Re: Shutdown tylan4 2002-05-24 09:59:58: needs test after TCA Clean finished:testrun completed; ok to use (gladys.snf.stanford.edu)
74: Problem tylan4 2002-07-05 11:35:21: TCACLEAN due to possible glass contamination (ychliu.snf.stanford.edu)
75: Re: Problem tylan4 2002-07-05 11:35:21: TCACLEAN due to possible glass contamination (seymour.snf.stanford.edu)
76: Problem tylan4 2002-08-08 11:51:29: Ox thickness 10% low... (latta.snf.stanford.edu)
77: Comment tylan4 2002-08-12 15:49:01: TYlan 4 (tberg.snf.stanford.edu)
78: Re: Problem tylan4 2002-08-08 11:51:29: Ox thickness 10% low...: testrun completed, thickness is good, ok to use (gladys.snf.stanford.edu)
79: Problem tylan4 2002-08-27 20:04:07: Intermittance H2 Fault (mcvittie.snf.stanford.edu)
80: Problem tylan4 2002-08-28 19:10:53: N2 Flow fault--Flow Now ok (mcvittie.snf.stanford.edu)
81: Use of Steam in Ox Furnaces (Mary Tang)
82: Problem tylan4 2002-08-28 19:28:31: Please do not use steam ox til further notice (mtang.snf.stanford.edu)
83: Shutdown tylan4 2002-09-03 09:48:41: N2 flow problem (tberg.snf.stanford.edu)
84: Re: Shutdown tylan4 2002-09-03 09:48:41: N2 flow problem (seymour.snf.stanford.edu)
85: Re: Problem tylan4 2002-08-28 19:10:53: N2 Flow fault--Flow Now ok (seymour.snf.stanford.edu)
86: Re: Problem tylan4 2002-08-27 20:04:07: Intermittance H2 Fault (seymour.snf.stanford.edu)
87: Problem tylan4 2002-09-04 15:41:14: see Ty #1 Equip email achieve note on TCA (seymour.snf.stanford.edu)
88: Shutdown tylan4 2002-09-05 06:38:26: H2 leak alarm (tberg.snf.stanford.edu)
89: Re: Shutdown tylan4 2002-09-05 06:38:26: H2 leak alarm (seymour.snf.stanford.edu)
90: Re: Problem tylan4 2002-08-28 19:28:31: Please do not use steam ox til further notice (seymour.snf.stanford.edu)
91: Shutdown tylan4 2002-09-06 08:48:08: doing a quick check (seymour.snf.stanford.edu)
92: Re: Shutdown tylan4 2002-09-06 08:48:08: doing a quick check (seymour.snf.stanford.edu)
93: Shutdown tylan4 2002-09-09 09:28:30: checking all gas lines to tube (seymour.snf.stanford.edu)
94: Re: Shutdown tylan4 2002-09-09 09:28:30: checking all gas lines to tube (seymour.snf.stanford.edu)
95: Shutdown tylan4 2002-09-24 15:58:30: needs uniformity test. (seymour.snf.stanford.edu)
96: Comment tylan4 2002-09-27 12:18:43: I ran the backup t/c calibration routine (seymour.snf.stanford.edu)
97: Re: Problem tylan4 2002-09-04 15:41:14: see Ty #1 Equip email achieve note on TCA (latta.snf.stanford.edu)
98: Re: Shutdown tylan4 2002-09-24 15:58:30: needs uniformity test. (latta.snf.stanford.edu)
99: Shutdown tylan4 2002-10-24 14:51:12: Pull and clean (mdickey.snf.stanford.edu)