Subject: Re: Problem wbsilicide SNF 2008-05-04 13:16:42: 4:1 piranha pot does not drain !
From: shott@snf.stanford.edu
Date: Sun, 4 May 2008 18:27:54 -0700

Ritesh:
The fitting that supplies the water that "powers" the aspirator had come out.  It is
a 1/4 inch teflon line with a FastNTite fitting ... which is the wrong fitting for this
job!!!!
I have replaced it and have emptied the hot pot.  However, it will continue to fail
until it is replaced with a proper fitting.
Thanks,
John