Messages 0xx
01:
Shutdown drytek1: shutdown test #2
(
(none)
)
02:
Shutdown drytek1: Test #2 Cleared
(
(none)
)
03:
Shutdown drytek1:
(
(none)
)
04:
Shutdown drytek1: cleared - shutdown because of blower
(
(none)
)
05:
Problem drytek1:
(
sigari.snf.stanford.edu
)
06:
Problem Clear drytek1: light bulb is bad, push botton switch is ok. changed bulb...
(
cbaxter.snf.stanford.edu
)
07:
Problem drytek1: bird meter not working
(
lindaw.snf.stanford.edu
)
08:
Problem drytek1: no plasma
(
ericp.snf.stanford.edu
)
09:
Problem Clear drytek1: slug was inserted all the way in..
(
cbaxter.snf.stanford.edu
)
10:
Problem Clear drytek1: ran recipe 3 and plasma was ok @100watts and 0 reflective.
(
cbaxter.snf.stanford.edu
)
11:
Problem drytek1:
(
sulchek.snf.stanford.edu
)
12:
Problem Clear drytek1: Turned pumps on. jim
(
mcvittie.snf.stanford.edu
)
13:
Shutdown drytek1: need to clean chamber electrode
(
cbaxter.snf.stanford.edu
)
14:
Shutdown Clear drytek1: cha. clean completed and condition..
(
cbaxter.snf.stanford.edu
)
15:
Shutdown drytek1: Vacum Problem
(
flannery.snf.stanford.edu
)
16:
Shutdown Clear drytek1: Turned on blower pump. Jim
(
mcvittie.snf.stanford.edu
)
17:
Problem drytek1: F115 tank too low
(
lindaw.snf.stanford.edu
)
18:
Problem Clear drytek1: F115 replenshed.
(
latta.snf.stanford.edu
)
19:
Shutdown drytek1: pump needed oil and filter changedDate: Nov 15, 2000 7:04:32 PM
(
cbaxter.snf.stanford.edu
)
20:
Shutdown drytek1: pump needed oil and filter changedDate: Nov 15, 2000 9:10:27 PM
(
cbaxter.snf.stanford.edu
)
21:
Problem drytek1 2001-01-09 09:56:58: Test problem ... please ignore
(
shott.snf.stanford.edu
)
22:
Re: Problem drytek1 2001-01-09 09:56:58: Test problem ... please ignore
(
shott.snf.stanford.edu
)
23:
Problem drytek1 2001-01-30 19:54:42: Will not start gas flow after starting process
(
sminne.snf.stanford.edu
)
24:
Re: Problem drytek1 2001-01-30 19:54:42: Will not start gas flow after starting process
(
cbaxter.snf.stanford.edu
)
25:
Problem drytek1 2001-02-28 11:30:19: can't tune
(
franklin.snf.stanford.edu
)
26:
Problem drytek1 2001-02-28 13:10:31: coil won't turn
(
lindaw.snf.stanford.edu
)
27:
Re: Problem drytek1 2001-02-28 13:10:31: coil won't turn
(
cbaxter.snf.stanford.edu
)
28:
Re: Problem drytek1 2001-02-28 11:30:19: can't tune
(
cbaxter.snf.stanford.edu
)
29:
Shutdown drytek1 2001-03-03 16:51:08: Tuning problem
(
mcp.snf.stanford.edu
)
30:
Re: Shutdown drytek1 2001-03-03 16:51:08: Tuning problem
(
cbaxter.snf.stanford.edu
)
31:
Problem drytek1 2001-03-17 20:03:08: gas flow could not be turned on
(
yangxu.snf.stanford.edu
)
32:
Re: Problem drytek1 2001-03-17 20:03:08: gas flow could not be turned on
(
cbaxter.snf.stanford.edu
)
33:
Problem drytek1 2001-03-21 15:51:50: Start cycle does not work
(
ganesan.snf.stanford.edu
)
34:
Re: Problem drytek1 2001-03-21 15:51:50: Start cycle does not work
(
cbaxter.snf.stanford.edu
)
35:
Shutdown drytek1 2001-04-17 10:20:24: Pumps down, but no gas flow on any process
(
lindaw.snf.stanford.edu
)
36:
Re: Shutdown drytek1 2001-04-17 10:20:24: Pumps down, but no gas flow on any process
(
booth.snf.stanford.edu
)
37:
Problem drytek1 2001-04-21 16:07:47: process 2 button lamp burnt out
(
lindaw.snf.stanford.edu
)
38:
Re: Problem drytek1 2001-04-21 16:07:47: process 2 button lamp burnt out
(
booth.snf.stanford.edu
)
39:
Shutdown drytek1 2001-04-24 07:46:08: vacuum system not reaching base pressure
(
booth.snf.stanford.edu
)
40:
Re: Shutdown drytek1 2001-04-24 07:46:08: vacuum system not reaching base pressure
(
cbaxter.snf.stanford.edu
)
41:
Shutdown drytek1 2001-05-14 22:01:15: inductor arcing failure
(
storment.snf.stanford.edu
)
42:
Problem drytek1 2001-05-14 23:05:21: adjustable inductor is arcing will need to change..
(
cbaxter.snf.stanford.edu
)
43:
Re: Shutdown drytek1 2001-05-14 22:01:15: inductor arcing failure
(
cbaxter.snf.stanford.edu
)
44:
Re: Problem drytek1 2001-05-14 23:05:21: adjustable inductor is arcing will need to change..
(
cbaxter.snf.stanford.edu
)
45:
Problem drytek1 2001-06-11 09:59:58: missing locating pin front right
(
lindaw.snf.stanford.edu
)
46:
Re: Problem drytek1 2001-06-11 09:59:58: missing locating pin front right
(
cbaxter.snf.stanford.edu
)
47:
Problem drytek1 2001-06-29 10:08:58: load complete light out
(
franklin.snf.stanford.edu
)
48:
Re: Problem drytek1 2001-06-29 10:08:58: load complete light out
(
cbaxter.snf.stanford.edu
)
49:
Shutdown drytek1 2001-07-18 10:31:49: chamber will not pump down to base press.
(
booth.snf.stanford.edu
)
50:
Shutdown drytek1 2001-07-18 11:51:17: Blower needs oil (as per Len)
(
latta.snf.stanford.edu
)
51:
Re: Shutdown drytek1 2001-07-18 11:51:17: Blower needs oil (as per Len)
(
cbaxter.snf.stanford.edu
)
52:
Re: Shutdown drytek1 2001-07-18 10:31:49: chamber will not pump down to base press.
(
cbaxter.snf.stanford.edu
)
53:
Re: Shutdown drytek1 2001-07-18 10:31:49: chamber will not pump down to base press.
(
cbaxter.snf.stanford.edu
)
54:
Shutdown drytek1 2001-08-25 07:44:16: facility
(
rcrane.snf.stanford.edu
)
55:
Shutdown drytek1 2001-08-25 07:44:46: facility
(
rcrane.snf.stanford.edu
)
56:
Re: Shutdown drytek1 2001-08-25 07:44:46: facility
(
rcrane.snf.stanford.edu
)
57:
Re: Shutdown drytek1 2001-08-25 07:44:16: facility
(
rcrane.snf.stanford.edu
)
58:
Problem drytek1 2001-08-26 20:33:22: Teflon pegs
(
dton.snf.stanford.edu
)
59:
Re: Problem drytek1 2001-08-26 20:33:22: Teflon pegs
(
cbaxter.snf.stanford.edu
)
60:
Problem drytek1 2001-08-29 22:34:48: etch rate unusually low; spoke to others who observed same problem
(
sjlee.snf.stanford.edu
)
61:
Problem drytek1 2001-08-31 18:59:09: Slow etch rate - reconfirmed
(
dton.snf.stanford.edu
)
62:
Re: Problem drytek1 2001-08-31 18:59:09: Slow etch rate - reconfirmed
(
cbaxter.snf.stanford.edu
)
63:
Re: Problem drytek1 2001-08-29 22:34:48: etch rate unusually low; spoke to others who observed same problem
(
cbaxter.snf.stanford.edu
)
64:
Shutdown drytek1 2001-09-04 13:43:05: checking for water leaks in chamber.
(
booth.snf.stanford.edu
)
65:
Shutdown drytek1 2001-09-04 22:13:27:
(
cbaxter.snf.stanford.edu
)
66:
Re: Shutdown drytek1 2001-09-04 22:13:27:
(
cbaxter.snf.stanford.edu
)
67:
Re: Shutdown drytek1 2001-09-04 13:43:05: checking for water leaks in chamber.
(
cbaxter.snf.stanford.edu
)
68:
Problem drytek1 2001-09-10 22:17:00:
(
crozier.snf.stanford.edu
)
69:
Re: Problem drytek1 2001-09-10 22:17:00:
(
cbaxter.snf.stanford.edu
)
70:
Shutdown drytek1 2001-09-13 13:41:52: system won't pumpdown
(
booth.snf.stanford.edu
)
71:
Re: Shutdown drytek1 2001-09-13 13:41:52: system won't pumpdown
(
booth.snf.stanford.edu
)
72:
Shutdown drytek1 2001-09-17 07:14:59: short pump maintenance.
(
booth.snf.stanford.edu
)
73:
Re: Shutdown drytek1 2001-09-17 07:14:59: short pump maintenance.
(
booth.snf.stanford.edu
)
74:
Shutdown drytek1 2001-09-18 06:54:16: short pump adjustment.
(
booth.snf.stanford.edu
)
75:
Re: Shutdown drytek1 2001-09-18 06:54:16: short pump adjustment.
(
booth.snf.stanford.edu
)
76:
Problem drytek1 2001-10-18 16:15:30: No Plasma
(
florando.snf.stanford.edu
)
77:
Re: Problem drytek1 2001-10-18 16:15:30: No Plasma
(
cbaxter.snf.stanford.edu
)
78:
Comment drytek1 2001-10-23 11:25:01: No Plasma
(
florando.snf.stanford.edu
)
79:
Shutdown drytek1 2001-10-23 11:29:26: NO plasma
(
florando.snf.stanford.edu
)
80:
Re: Shutdown drytek1 2001-10-23 11:29:26: Fixed by Increase The RF Level
(
hphan.snf.stanford.edu
)
81:
Problem drytek1 2001-11-12 10:32:29: Mechanical Pump Light is Off, Cycle Doesn't Work
(
hphan.snf.stanford.edu
)
82:
Re: Problem drytek1 2001-11-12 10:32:29: Mechanical Pump Light is Off, Cycle Doesn't Work
(
booth.snf.stanford.edu
)
83:
Shutdown drytek1 2001-12-15 00:00:03: no plasma
(
hector.snf.stanford.edu
)
84:
Re: Shutdown drytek1 2001-12-15 00:00:03: Re-tuned Matching Network
(
mcvittie.snf.stanford.edu
)
85:
Problem drytek1 2001-12-17 11:22:12: Can't adjust the coil. -> High reflectance power
(
pzwahlen.snf.stanford.edu
)
86:
Re: Problem drytek1 2001-12-17 11:22:12: Can't adjust the coil. -> High reflectance power
(
booth.snf.stanford.edu
)
87:
Problem drytek1 2001-12-17 19:24:48: plasma stops
(
rgrow.snf.stanford.edu
)
88:
Re: Problem drytek1 2001-12-17 19:24:48: plasma stops
(
cbaxter.snf.stanford.edu
)
89:
Problem drytek1 2001-12-18 20:35:49: intermitent plasma
(
hector.snf.stanford.edu
)
90:
Re: Problem drytek1 2001-12-18 20:35:49: intermitent plasma
(
booth.snf.stanford.edu
)
91:
Shutdown drytek1 2002-01-07 16:39:31: do for oil change
(
cbaxter.snf.stanford.edu
)
92:
Re: Shutdown drytek1 2002-01-07 16:39:31: do for oil change
(
cbaxter.snf.stanford.edu
)
93:
Shutdown drytek1 2002-02-14 16:13:00: need oil change
(
cbaxter.snf.stanford.edu
)
94:
Re: Shutdown drytek1 2002-02-14 16:13:00: need oil change
(
cbaxter.snf.stanford.edu
)
95:
Problem drytek1 2002-03-16 14:29:41: drytek1 won't pump down
(
jackson.snf.stanford.edu
)
96:
Shutdown drytek1 2002-03-18 11:51:53: Throtle valve controller out of calibration
(
hphan.snf.stanford.edu
)
97:
Re: Shutdown drytek1 2002-03-18 11:51:53: Throtle valve controller out of calibration
(
cbaxter.snf.stanford.edu
)
98:
Re: Problem drytek1 2002-03-16 14:29:41: drytek1 won't pump down
(
cbaxter.snf.stanford.edu
)
99:
Shutdown drytek1 2002-03-19 18:36:29: shutdown for oil and filter change
(
cbaxter.snf.stanford.edu
)
[<<-]
[<-]
[->]
[->>]
|