Subject: raith issues?
From: Charis Quay Huei Li <chquay@MtHolyoke.edu>
Date: Sat, 7 Aug 2004 04:01:06 -0400 (EDT)
(1) beam drift/settling right after beam is unblanked, same as has been
seen before. several hundred nm.
(2) stigmation varies from place to place on sample by several %.
(3) focus also drifts over time.
problem (3) could be due to the fact that i'm using carbon dots but i've
bever seen (2) when i'm not near a clip. what was the problem the last
time we saw (1)?
charis.