Subject: problem report
From: Takashi Kodama <kodama@stanford.edu>
Date: Thu, 5 May 2011 18:07:57 -0700 (PDT)


Hi all,

The pop-up error message "pattern generator is not ready, demo mode enabled" was appeared
when I started the raith software. According to the James's instruction, I could solve
the problem.
Please follow the way if you have same issue.
1. shutdown the raith PC
2. turn off the hardware (only second controller "HIGH PERFORMANCE SCAN PROCESSOR" from
the top).
3. turn on it
4. restart PC

Thanks,

Takashi Kodama