Greetings:
System is back up and we will be testing it through this afternoon.
We had experienced several errors over the weekend that appeared to be a loss of the IEE-488 GPIB communication bus that controls the XYZT stage controller, the piezo controls and the picoammeter.  These were reestablished and reset very early Sunday Morning.

These errors too were new to us thus far this week, and we will be testing and observing the system carefully.

Thank you,

James W. Conway

-------- Original Message --------
Subject: Re: Problem raith SNF 2005-02-15 02:38:08: stage stuck in load lock procedure
Date: Tue., 15 Feb. 2005 11:46:10 -0800
From: jwc@snf.stanford.edu
To: raith-pcs@snf.stanford.edu


Cleared:
Emergency error no 2:  PRESUMABLY communication failure from transfer arm inner limit switch .
emergency error level 6:  presumably time out on turbo pump.  May have been a previous turbo time out or not reach vacuum set point error.  Could not find in log files.
JWC 02152005:11:39:00