Subject: Changes observed in characteristic of the focus and stigmation controls on the RAITH , User please report if they observed changes in control functions.
From: James Conway <jwc@snf.stanford.edu>
Date: Wed, 14 Jun 2006 18:39:10 -0700
Wed, 14 Jun 2006 18:39:10 -0700

Greetings Raith Community:

Today when I came onto the system it was immediately apparent that the 
typical response characteristic  manipulating the controls for FOCUS and 
less so the Stigmation controls has changed profoundly from my last 
access on the system yesterday.  In fact I have never seen it take so 
much translation of the controls to change focus and orders of magnitude 
more current is required to properly stigmate. (Note before that it was 
less than 1 mm to go from stop to stop on the X-Y stigmation controls. 
now it is five to ten mm!)

 This indicated to me that the chamber and likely the lower end of the 
column at the objective lens and pole piece have become severely 
contaminated in an individual event.  The upper column and gun section 
seems to be maintaining good vacuum, but the big change in stigmation 
characteristic is very worrisome.

I suspect that the RAITH system may have become contaminated with the 
oily residues that we have been observing showing up on the VINOX 
microscope.  See my previous message on this topic. 
<http://snf.stanford.edu/cgi-bin/ezmlm-cgi?250:mss:768:200606:eiilfbknkjdbobaehjam>

In fact it may be an Ebeam Lab Raith User whom  introduced oil or 
silicon contaminated samples into the system yesterday evening.

All Users working on the tool are asked to be especially vigilant in not 
placing samples into the system that are not fully baked and cured to 
above 140 degrees C.  Never allow any materials into the system 
containing Wax, greases, silicon based lubricants and pump oils, krytox 
fluoridated grease, adhesive tapes including carbon and copper tape.  If 
it isn't high vacuum compliant  it doesn't go into the system.  If 
residues are on the backside of the sample it doesn't go into the system.

If it is not one of our standard resist materials it is not to go into 
the system without my approval.

In the past on other SEM and Ebeam systems, we have observed spurious 
contamination events from out gassing of resist and extended exposure of 
resist under the Ebeam.  In time this will coat up the Apertures, the 
pole pieces, and the Objective lens with hydrocarbon residue resulting 
in more and more current required to the properly align the gun and when 
focusing and stigmating on a SEM image.  This is what I would normally 
observe as a SEM system extend beyond the normal PM interval for 
cleaning, FE gun and aperture replacement... Some of those events were 
recoverable and the systems returned to normal after a few days period 
of time. But most of the time contamination like this can only be 
addressed by a full column PM and an extended N2 Purge followed by a 
several hundred hour long Gun Bake Out.

In reference to the superb resolution and outstanding EBL results we are 
obtaining on the system since a new column was installed on this tool, 
it would be unfortunate if just one careless user ruined the whole thing 
for everyone by contaminating the column.  I am hoping we will not have 
to tear this column down to effect a repair.  Users are requested to 
report any changes they have observed when running the system recently.

Thank you for your support!

James Conway







Greetings Raith Community:

Today when I came onto the system it was immediately apparent that the typical response characteristic  manipulating the controls for FOCUS and less so the Stigmation controls has changed profoundly from my last access on the system yesterday.  In fact I have never seen it take so much translation of the controls to change focus and orders of magnitude more current is required to properly stigmate. (Note before that it was less than 1 mm to go from stop to stop on the X-Y stigmation controls. now it is five to ten mm!)

 This indicated to me that the chamber and likely the lower end of the column at the objective lens and pole piece have become severely contaminated in an individual event.  The upper column and gun section seems to be maintaining good vacuum, but the big change in stigmation characteristic is very worrisome.

I suspect that the RAITH system may have become contaminated with the oily residues that we have been observing showing up on the VINOX microscope.  See my previous message on this topic. In fact it may be an Ebeam Lab Raith User whom  introduced oil or silicon contaminated samples into the system yesterday evening.

All Users working on the tool are asked to be especially vigilant in not placing samples into the system that are not fully baked and cured to above 140 degrees C.  Never allow any materials into the system containing Wax, greases, silicon based lubricants and pump oils, krytox fluoridated grease, adhesive tapes including carbon and copper tape.  If it isn't high vacuum compliant  it doesn't go into the system.  If residues are on the backside of the sample it doesn't go into the system.

If it is not one of our standard resist materials it is not to go into the system without my approval.

In the past on other SEM and Ebeam systems, we have observed spurious contamination events from out gassing of resist and extended exposure of resist under the Ebeam.  In time this will coat up the Apertures, the pole pieces, and the Objective lens with hydrocarbon residue resulting in more and more current required to the properly align the gun and when focusing and stigmating on a SEM image.  This is what I would normally observe as a SEM system extend beyond the normal PM interval for cleaning, FE gun and aperture replacement... Some of those events were recoverable and the systems returned to normal after a few days period of time. But most of the time contamination like this can only be addressed by a full column PM and an extended N2 Purge followed by a several hundred hour long Gun Bake Out.

In reference to the superb resolution and outstanding EBL results we are obtaining on the system since a new column was installed on this tool, it would be unfortunate if just one careless user ruined the whole thing for everyone by contaminating the column.  I am hoping we will not have to tear this column down to effect a repair.  Users are requested to report any changes they have observed when running the system recently.

Thank you for your support!

James Conway