From: James Conway <jwc@snf.stanford.edu>
Date: Mon, 02 Apr 2007 14:13:25 -0700

RAITH Users:

Measurements I have been making of each landing of standard sample holder onto the piezo mounts show that we did not fully correct the problems during our service work two weeks ago. We are continuing to make measurements and seeking an effective solution.
System will continue in yellow light mode. 

Users reports of all write results are requested to this email list.

Thank you,

James Conway

-------- Original Message --------
Subject: Problem raith SNF 2007-03-30 10:24:33: Stage piezo post heights are different cannot level samples
Date: Fri, 30 Mar 2007 10:24:34 -0700
From: jwc@snf.stanford.edu
To: raith-pcs@snf.stanford.edu


Stage Piezo post height P1 is ~ 245 um too high resulting in the standard sample holder not landing flat onto the X_Y stage.
We are working to resolve this situation working with RAITH Field service support.
Users are recommended to layout their patterns along the U axis and stay within a 10 mm V area or re focus and align write field alignment every 10 mm in V. 
Stitching and overlay registration is out of specification.

JWC