Subject: Raith available till 10:00am
From: linyouc@stanford.edu
Date: Wed, 28 May 2008 09:19:02 -0700

Done earlier. It seems to work very well so far.

Linyou

Quoting Byoungil Lee <bilee@stanford.edu>:

> It's a problem caused by a switch in the controller box.
> The switch called 'ebeam' should be up position, but sometimes it's
> mysteriously switched down...
> It was okay until 3am today.
>
> Byoungil
>
> ----- Original Message ----- From: "Salman Latif" <slatif@stanford.edu>
> To: "Raith SNF Mailing list" <raith@snf.stanford.edu>
> Sent: Wednesday, May 28, 2008 5:15 AM
> Subject: finishing early -- beam blanking issues
>
>
>> The beam blank button seemed to be in reverse mode. I could see the  
>>   SEM image in the off mode, and vice versa. More importantly, the   
>> beam  would not stay blanked during a move from one position to   
>> another  (whether you turn the blanker on or off). I shut down both  
>>  computers,  and after starting up again, could not see the SEM   
>> image at all.
>>



-- 
Linyou Cao
PhD candidate
Dept. of Mater. Sci.and Engin., Brongersma Group
President&Co-founder
Stanford Nanoscience&Nanotechnology Society
Stanford University