Subject: Re: Raith update
From: toecutter4ranger@gmail.com
Date: Sun, 1 Jun 2008 13:32:43 +0000

 Sun, 1 Jun 2008 13:32:43 +0000
Just use the tool and send me email a d be sure to log your sessions in the logbook.

Jwc
Sent via BlackBerry by AT&T

-----Original Message-----
From: "Byoungil Lee" <bilee@stanford.edu>

Date: Sun, 1 Jun 2008 00:39:05 
To:"Jesse Lu" <jesselu@stanford.edu>
Cc:"John Shott" <shott@stanford.edu>,"Raith SNF Mailing list" <raith@snf.stanford.edu>,<toecutter4ranger@gmail.com>
Subject: Re: Raith update


Dear All, 
  
I thought I was going to be able to use the tool. But, since the tool is still in red,
I assume I do not have access to the tool.  (or, can I try to use it? ) 
I am releasing my session from 0am to 4am. 
  
  
  
Thank you, 
Byoungil 
  
----- Original Message ----- 
From: Xinran 
  Wang <mailto:xinranw@stanford.edu>  
To: Jesse Lu <mailto:jesselu@stanford.edu>  
Cc: John Shott <mailto:shott@stanford.edu>  ; Raith SNF 
  Mailing list <mailto:raith@snf.stanford.edu>  ; toecutter4ranger@gmail.com <mailto:toecutter4ranger@gmail.com>
 
Sent: Saturday, May 31, 2008 10:58 PM 
Subject: Re: Raith update 

Hi Jesse,
Thank you for your help to bring raith up. But raith is still in red light mode in coral,
if this is the case, can we still use it or someone has to turn it to yellow first?


Thanks,
Xinran


 
On Sat, May 31, 2008 at 2:48 PM, Jesse Lu <jesselu@stanford.edu <mailto:jesselu@stanford.edu>
> wrote:
 Hey Raith users,

After a few hours on the phone with James and a Raith representative, we seem to be
able to load samples now. The T translation system seems to be still acting up, but
we were able to successfully load a sample this afternoon at 2pm. However, we used a
modified loading procedure. 

James says to: use a modified load procedure
 
 
* Do not simply hit the Load Sample button in the Loadlock window to drive your sample
into the e-beam chamber. 
* Instead, perform a Find Home position for T procedure 
 
* This is done by first going to the Find Home window

 
 
* Then click on the lightning rod button on the T row 
* Wait a few seconds for the procedure to finish 
* Finally, check the Delta value for T in the Find Home window 
* Please perform this procedure 2-3 times until the Delta value is on the order of 20-50
microns
 
* Then proceed with the normal load procedure by clicking the Load Sample button in
the Loadlock window 
John Shott, James would also like you to put the system back in yellow light mode if
possible.

Finally, users Mitsuru and Kathryn are currently "sharing the ride" on the tool. Let's
revert back to the usual schedule at midnight tonight with Bilee's reservation. If extra
time is needed because of this delay, let's be flexible with one another.

Jesse