["multipart/related" not shown]
Fri, 31 Oct 2008 15:55:49 -0800
Subject: Raith status Friday afternoon -=- System is down and offline to all
users this weekend. Expect to be back up for user next Tuesday afternoon.
From: James Conway <jwc@snf.stanford.edu>
Date: Fri, 31 Oct 2008 15:55:49 -0800
Greetings,
The Raith system has serious problems in the Stage controller
electronics modules resulting in virtual increases in the X, Y, Z, T
Stage coordinates.
It is unsuitable for lithography and currently we cannot complete the
Load Lock sequence without errors.
No viruses were found on the RAITH or off-line Computers during sweeps
of these two systems yesterday.
Best expectation when we will be up and qualified for Users is sometime
Tuesday afternoon next week.
We are waiting for several motor controller PC boards to shipped to us
from NY State.
All the Very Best for a Happy Halloween Holiday,
James Conway

Fri, 31 Oct 2008 15:55:49 -0800
Subject: Raith status Friday afternoon -=- System is down and offline to all
users this weekend. Expect to be back up for user next Tuesday afternoon.
From: James Conway <jwc@snf.stanford.edu>
Date: Fri, 31 Oct 2008 15:55:49 -0800
["image/gif" not shown]
Fri, 31 Oct 2008 15:55:49 -0800