Subject: Re: problem on hardware communication
From: Krishna Coimbatore Balram <kcbalram@stanford.edu>
Date: Wed, 31 Aug 2011 03:26:46 -0700 (PDT)

Hi Takashi,

I also got the same error. The reason why that showed up was that the smartSEM was not
open in remcon mode.

I exited the smart sem and restarted in the remcon mode and the system works fine.

Best,
Krishna

----- Original Message -----
From: "Takashi Kodama" <kodama@stanford.edu>
To: raith@snf.stanford.edu
Sent: Saturday, August 27, 2011 5:01:05 PM
Subject: problem on hardware communication


Hi all,

When I launched the raith software, the message was appeared on raith software.

communication with the column control computer not possible
please check the connection and retry

I hardly operate the system so I release my reservation till 9:30 pm.
I am not sure it this is common issue for all member or it comes from my account setting
or not.

Thanks,

Takashi Kodama