Subject: New RAITH User Logins, Please check your magnification display setting on LEO.
From: James Conway <jwc@snf.stanford.edu>
Date: Wed, 11 Jun 2003 13:56:23 -0700

 
Hello Raith Users:

New Users Logins:
All those Users whom received new logins in the last month may wish to verify that their user accounts are using the correct Magnification Display Output device on the LEO SEM configuration set up page.

*******************
IF YOU HAVE EXPERIENCED DIFFICULTIES DURING WRITE FIELD ALIGNMENT ...
*******************

Symptoms:
When during the initial Write Field alignment procedure where on the initial image acquisition the particle is not found anywhere in the acquired image field of view, please check the Magnification Display Output device on the LEO 'SEM CONDITIONS' set up

Adjustments:
This can be checked and adjusted as follows:
 

This ensures that the correct Magnification Lookup Table is used by both the RAITH and LEO sides of the system during image acquisitions during your write field alignment procedure.

Please see James Conway if you need help with checking or changing these settings.

Thank you,

James Conway
E-beam Technology Group.
Stanford Nanofabrication Facility
650-725-7075