[Messages by date]
[SNF Home]
[SNF Archive Home]
[Subscribe To List]
[List FAQ]
Threads for December 2006
December 2006
[1] Comment raith SNF 2006-12-01 13:44:43: Very nice beam waist today at 10 kV 30 um aperture
[1] Re: Comment raith SNF 2006-10-27 13:18:48: User can now operate system to not to exceed 20 kV
[1] Re: Comment raith SNF 2006-10-27 13:21:44: Piezo leveling post havew been calibrated.
[1] Comment raith SNF 2006-12-01 13:45:26: Users can operate at 0.3 to 30 keV
[2] Re: Shutdown raith SNF 2006-12-08 02:03:00: SEM does not work.
[2] Re: Shutdown raith SNF 2006-12-08 11:17:26: E/O card or image acquisition on LEO faulty
[1] Raith Status: Monday December 11, 2006 4 PM -- RAITH WILL REMAIN OFFLINE FOR REPAIRS.
[2] Re: Shutdown raith SNF 2006-12-08 12:26:52: E/O card damaged in EHT arc event...
[2] Re: Shutdown raith SNF 2006-12-08 14:51:57: Electron optics board may have been damaged im EHT trip event.
[2] Re: Problem raith SNF 2006-12-08 10:33:58: Std sample holder was loaded improperly
[1] Comment raith SNF 2006-12-15 12:49:45: System qualification writing. All seems fine...
[1] Comment raith SNF 2006-12-18 08:09:59: LeoPC crashed during unload, LL error
[1] Shutdown raith SNF 2006-12-18 11:04:55: off line to users == holdiay shutdown and bake out
[<<-]
[<-]
[->]
[->>]