Messages 5xx


00: Comment raith SNF 2006-07-15 13:00:38: EHT trip (yiyang.snf.stanford.edu)
01: Problem raith SNF 2006-07-15 13:05:19: EHT trip (yiyang.snf.stanford.edu)
02: Comment raith SNF 2006-07-15 21:15:13: Load Lock Error (yjhuo.snf.stanford.edu)
03: Re: Problem raith SNF 2006-07-15 13:05:19: EHT trip (jwc.snf.stanford.edu)
04: Re: Comment raith SNF 2006-07-15 13:00:38: EHT trip (jwc.snf.stanford.edu)
05: Re: Comment raith SNF 2006-07-15 21:15:13: Load Lock Error (jwc.snf.stanford.edu)
06: Comment raith SNF 2006-07-18 16:55:08: System tested for operations at 5, 10, 15, 20, 25, and 30 kV (jwc.snf.stanford.edu)
07: Comment raith SNF 2006-07-28 11:18:11: 30 kV test completed Users can use any ACCELERATION Voltage 0.3 to 30 KV (jwc.snf.stanford.edu)
08: Problem raith SNF 2006-07-29 15:46:07: Fatal Error 5 (andrei.snf.stanford.edu)
09: Comment raith SNF 2006-07-29 22:16:24: Raith Error 5 cleared (jhemanth.snf.stanford.edu)
10: Re: Problem raith SNF 2006-07-29 15:46:07: Fatal Error 5 (jwc.snf.stanford.edu)
11: Re: Comment raith SNF 2006-07-29 22:16:24: Raith Error 5 cleared (jwc.snf.stanford.edu)
12: Re: Comment raith SNF 2006-07-07 14:13:56: System is being qualified for Users today (jwc.snf.stanford.edu)
13: Re: Comment raith SNF 2006-07-07 17:10:04: Raith is qualified (jwc.snf.stanford.edu)
14: Re: Comment raith SNF 2006-07-14 13:05:56: System tested to 15 kV today. Users restricted to BELOW 15 KV ACC. VOLTAGE (jwc.snf.stanford.edu)
15: Problem raith SNF 2006-08-11 01:11:58: (englund.snf.stanford.edu)
16: Shutdown raith SNF 2006-08-11 01:12:04: (englund.snf.stanford.edu)
17: Shutdown raith SNF 2006-08-11 01:12:26: load lock error (englund.snf.stanford.edu)
18: Re: Shutdown raith SNF 2006-08-11 01:12:04: (jwc.snf.stanford.edu)
19: Re: Problem raith SNF 2006-08-11 01:11:58: (jwc.snf.stanford.edu)
20: Re: Shutdown raith SNF 2006-08-11 01:12:26: load lock error (jwc.snf.stanford.edu)
21: Problem raith SNF 2006-08-11 10:25:10: nested load lock errors and a manual vent of system chamber (jwc.snf.stanford.edu)
22: Comment raith SNF 2006-08-11 11:27:03: users queue will resume at 3 - 4 PM (jwc.snf.stanford.edu)
23: Re: Problem raith SNF 2006-08-11 10:25:10: nested load lock errors and a manual vent of system chamber (jwc.snf.stanford.edu)
24: Re: Comment raith SNF 2006-08-11 11:27:03: users queue will resume at 3 - 4 PM (jwc.snf.stanford.edu)
25: Re: Comment raith SNF 2006-07-18 16:55:08: System tested for operations at 5, 10, 15, 20, 25, and 30 kV (jwc.snf.stanford.edu)
26: Teacher, Student, School - Academic Software Discounts (Educational Sales Teams)
27: Problem raith SNF 2006-08-22 00:56:40: ULD failed (lorry.snf.stanford.edu)
28: Re: Problem raith SNF 2006-08-22 00:56:40: ULD failed (jwc.snf.stanford.edu)
29: Problem raith SNF 2006-08-26 11:00:34: Beam current measurement (rohank.snf.stanford.edu)
30: Shutdown raith SNF 2006-09-06 15:30:25: power off (gyzhang.snf.stanford.edu)
31: Re: Problem raith SNF 2006-08-26 11:00:34: Beam current measurement (rcrane.snf.stanford.edu)
32: Comment raith SNF 2006-09-07 15:20:04: System on and stabilizing (rcrane.snf.stanford.edu)
33: Re: Shutdown raith SNF 2006-09-06 15:30:25: power off (rcrane.snf.stanford.edu)
34: Problem raith SNF 2006-09-08 08:16:59: Needs superuser checkout (rcrane.snf.stanford.edu)
35: Shutdown raith SNF 2006-09-08 16:37:43: sample can't be unloaded (gyzhang.snf.stanford.edu)
36: Re: Comment raith SNF 2006-07-28 11:18:11: 30 kV test completed Users can use any ACCELERATION Voltage 0.3 to 30 KV (jwc.snf.stanford.edu)
37: Re: Comment raith SNF 2006-09-07 15:20:04: System on and stabilizing (jwc.snf.stanford.edu)
38: Re: Problem raith SNF 2006-09-08 08:16:59: Needs superuser checkout (jwc.snf.stanford.edu)
39: Re: Shutdown raith SNF 2006-09-08 16:37:43: sample can't be unloaded (jwc.snf.stanford.edu)
40: Problem raith SNF 2006-09-11 11:02:12: system check out in progress .. coral schedule resumes this afternoon. (jwc.snf.stanford.edu)
41: Re: Problem raith SNF 2006-09-11 11:02:12: system check out in progress .. coral schedule resumes this afternoon. (jwc.snf.stanford.edu)
42: Comment raith SNF 2006-09-12 17:41:12: Users requested to report their EBL results to raith@snf.stanford.edu (jwc.snf.stanford.edu)
43: Comment raith SNF 2006-09-13 17:34:21: excellent stitching and resolution on qualification test 9112006 (jwc.snf.stanford.edu)
44: Comment raith SNF 2006-09-15 10:25:03: 120 um Gun alignment out of specification. (jwc.snf.stanford.edu)
45: Problem raith SNF 2006-09-19 09:29:10: Raith right PC (sanaka.snf.stanford.edu)
46: Re: Problem raith SNF 2006-09-19 09:29:10: Raith right PC (jwc.snf.stanford.edu)
47: Problem raith SNF 2006-09-19 12:20:06: transient signals to Column Chamber Valve and problems measuring pico ammeter (jwc.snf.stanford.edu)
48: Shutdown raith SNF 2006-09-19 14:56:25: cooling water issues causing Gun trips (jwc.snf.stanford.edu)
49: Shutdown raith SNF 2006-09-19 15:55:35: elevated gun pressures after water flow failure (jwc.snf.stanford.edu)
50: Re: Problem raith SNF 2006-09-19 12:20:06: transient signals to Column Chamber Valve and problems measuring pico ammeter (jwc.snf.stanford.edu)
51: Shutdown raith SNF 2006-09-19 16:06:22: RAITH System off line for BAKE OUT OF GUN (jwc.snf.stanford.edu)
52: Comment raith SNF 2006-10-12 12:10:40: system in qualification and system test (jwc.snf.stanford.edu)
53: Re: Comment raith SNF 2006-09-12 17:41:12: Users requested to report their EBL results to raith@snf.stanford.edu (jwc.snf.stanford.edu)
54: Re: Comment raith SNF 2006-09-13 17:34:21: excellent stitching and resolution on qualification test 9112006 (jwc.snf.stanford.edu)
55: Re: Comment raith SNF 2006-09-15 10:25:03: 120 um Gun alignment out of specification. (jwc.snf.stanford.edu)
56: Re: Shutdown raith SNF 2006-09-19 14:56:25: cooling water issues causing Gun trips (jwc.snf.stanford.edu)
57: Re: Shutdown raith SNF 2006-09-19 15:55:35: elevated gun pressures after water flow failure (jwc.snf.stanford.edu)
58: Re: Shutdown raith SNF 2006-09-19 16:06:22: RAITH System off line for BAKE OUT OF GUN (jwc.snf.stanford.edu)
59: Comment raith SNF 2006-10-12 17:22:47: Inital phase of qualification complete. RAITH Returns to service (jwc.snf.stanford.edu)
60: Problem raith SNF 2006-10-16 13:04:19: LEO 32 IMAGING SW MODULE CORRUPTED (koseki.snf.stanford.edu)
61: Re: Comment raith SNF 2006-10-12 12:10:40: system in qualification and system test (jwc.snf.stanford.edu)
62: Re: Comment raith SNF 2006-10-12 17:22:47: Inital phase of qualification complete. RAITH Returns to service (jwc.snf.stanford.edu)
63: Re: Problem raith SNF 2006-10-16 13:04:19: LEO 32 IMAGING SW MODULE CORRUPTED (jwc.snf.stanford.edu)
64: Shutdown raith SNF 2006-10-17 12:07:36: New computer does not hav all confguration and drivers complete on the PC (jwc.snf.stanford.edu)
65: Re: Shutdown raith SNF 2006-10-17 12:07:36: New computer does not hav all confguration and drivers complete on the PC (jwc.snf.stanford.edu)
66: Problem raith SNF 2006-10-17 17:24:50: Piezo and height control modules not enabled -- OK to write by most users (jwc.snf.stanford.edu)
67: Adobe, Microsoft, Symantec @ up to 70% OFF for Education (Education Computer Producst)
68: Problem raith SNF 2006-10-18 13:28:15: please report immediately if you smell smoke or electrical smell in Ebeam Lab (jwc.snf.stanford.edu)
69: Re: Problem raith SNF 2006-10-17 17:24:50: Piezo and height control modules not enabled -- OK to write by most users (jwc.snf.stanford.edu)
70: Problem raith SNF 2006-10-18 16:12:12: patch installation corrupted ebmcurrent.dll (jwc.snf.stanford.edu)
71: Re: Problem raith SNF 2006-10-18 13:28:15: please report immediately if you smell smoke or electrical smell in Ebeam Lab (jwc.snf.stanford.edu)
72: Re: Problem raith SNF 2006-10-18 16:12:12: patch installation corrupted ebmcurrent.dll (jwc.snf.stanford.edu)
73: Comment raith SNF 2006-10-19 18:25:29: system is up and functional NOT TO EXCEED 10 KV (jwc.snf.stanford.edu)
74: Comment raith SNF 2006-10-21 03:57:09: unloading error (hideoiw.snf.stanford.edu)
75: Problem raith SNF 2006-10-21 04:06:01: Unloading error (hideoiw.snf.stanford.edu)
76: Comment raith SNF 2006-10-21 10:39:46: sample unloaded (gyzhang.snf.stanford.edu)
77: Re: Comment raith SNF 2006-10-21 03:57:09: unloading error (jwc.snf.stanford.edu)
78: Re: Problem raith SNF 2006-10-21 04:06:01: Unloading error (jwc.snf.stanford.edu)
79: Re: Comment raith SNF 2006-10-21 10:39:46: sample unloaded (jwc.snf.stanford.edu)
80: Problem raith SNF 2006-10-26 01:11:21: Raith Application won't open (dinhthuc.snf.stanford.edu)
81: Re: Problem raith SNF 2006-10-26 01:11:21: Raith Application won't open (jwc.snf.stanford.edu)
82: Comment raith SNF 2006-10-27 13:18:48: User can now operate system to not to exceed 20 kV (jwc.snf.stanford.edu)
83: Comment raith SNF 2006-10-27 13:21:44: Piezo leveling post havew been calibrated. (jwc.snf.stanford.edu)
84: Problem raith SNF 2006-10-29 15:56:10: Load lock error (kocabas.snf.stanford.edu)
85: Re: Problem raith SNF 2006-10-29 15:56:10: Load lock error (jwc.snf.stanford.edu)
86: Re: Comment raith SNF 2006-10-19 18:25:29: system is up and functional NOT TO EXCEED 10 KV (jwc.snf.stanford.edu)
87: Problem raith SNF 2006-11-06 08:54:45: turbo pump in standby mode (panjun.snf.stanford.edu)
88: Re: Problem raith SNF 2006-11-06 08:54:45: turbo pump in standby mode (jwc.snf.stanford.edu)
89: Autodesk AutoCAD for Faculty & Students (AutoCAD Educational Sales at CPE)
90: Problem raith SNF 2006-11-27 18:05:38: LEO computer does not display images (makarova.snf.stanford.edu)
91: Re: Problem raith SNF 2006-11-27 18:05:38: LEO computer does not display images (jwc.snf.stanford.edu)
92: Comment raith SNF 2006-12-01 13:44:43: Very nice beam waist today at 10 kV 30 um aperture (jwc.snf.stanford.edu)
93: Re: Comment raith SNF 2006-10-27 13:18:48: User can now operate system to not to exceed 20 kV (jwc.snf.stanford.edu)
94: Re: Comment raith SNF 2006-10-27 13:21:44: Piezo leveling post havew been calibrated. (jwc.snf.stanford.edu)
95: Comment raith SNF 2006-12-01 13:45:26: Users can operate at 0.3 to 30 keV (jwc.snf.stanford.edu)
96: Shutdown raith SNF 2006-12-08 02:03:00: SEM does not work. (hideoiw.snf.stanford.edu)
97: Problem raith SNF 2006-12-08 10:33:58: Std sample holder was loaded improperly (jwc.snf.stanford.edu)
98: Re: Shutdown raith SNF 2006-12-08 02:03:00: SEM does not work. (jwc.snf.stanford.edu)
99: Shutdown raith SNF 2006-12-08 11:17:26: E/O card or image acquisition on LEO faulty (jwc.snf.stanford.edu)