Subject: Re: Problem raith SNF 2009-01-07 16:55:00: system qualification FAILED == poor write field to write field stitching
From: jwc@snf.stanford.edu
Date: Thu, 8 Jan 2009 17:31:24 -0800

Adjusted the power suopply voltage to the rate servo card and checked there voltages
to specification on the X motor controller == excellent stitching was obtainedon the
second exposures.  Overlay testing tomorrow in preparation for pattern transfer