Subject: Comment raith SNF 2012-08-25 17:30:58: System good to go
From: kcbalram@snf.stanford.edu

The high speed pattern processor was off.... turned it on and restarted the system...
works ok now
For future reference:
While logging in on the Raith PC, if you get a popup saying pattern generator off Demo
mode enabled, make sure the high speed pattern generator (second box from top in the
rack ) is turned on. You will need to do a hardware reboot afterwards (shutdown and
then turn on, just hitting restart may or may not work)
Best,
Krishna