Subject: Problem raith SNF 2004-09-30 11:49:30: raith Blanker leaking
From: jwc@snf.stanford.edu
Date: Thu, 30 Sep 2004 11:49:31 -0700

Raith blanker has been measured leaking:
Both blankers ON 0.0005 - 0.0007 nA
LEO blanker ON Raith Blanker OFF  0.0005 - 0.0008 nA
LEO Blanker OFF RAITH Blanker ON 0.0017 - 0.0018 nA
QED LEO Blanker is OK , RAITH Blanker is leaking.
RAITH has been called for service to this section.
JWC 1145 hours 9302004