Subject: Re: Problem raith SNF 2005-06-29 22:49:56: unable to image
From: jwc@snf.stanford.edu
Date: Thu, 30 Jun 2005 11:56:17 -0700

cleared 
System functionality restored...
At 22:15 06292005 there was an EHT trip at 10 kV on FUSHMAN login.
The L-REM module on LEO was lost or corrupted eliminating image acquisition SW on thne
platform.
Reseting hardware after SHUTTING DOWN AND RESTARTING both LEO and RAITH computers corrected
the problem.