Subject: Re: Problem raith SNF 2006-01-09 12:34:11: Zeiss Field service did not perform requested calibrations
From: jwc@snf.stanford.edu
Date: Mon, 9 Jan 2006 14:56:52 -0800

UPDATE from previous message at 12:30 today:
Evidently Zeiss SMT brought system up 1-2-06 and performed initial calibrations requested
but did not return the following day after 24 hours or more of FE-gun operation run
time.  Shifts in the magnification amplifier centering functions, and some non-linearity
as evidenced by shifting in image in the mag range three. 
Will follow up with Raith and Zeiss SMT Field service tomorrow morning to schedule for
more adjustments hopefully later this week.
JWC 14:53:00 01092006