Subject: Re: Shutdown raith SNF 2006-05-09 11:33:40: motor z or virtual control errors yet to be resolved.
From: jwc@snf.stanford.edu
Date: Tue, 9 May 2006 15:26:29 -0700

Troubleshooted Z motor on Stage and Virtual Motor control module.  Reduced the speed
of the Motor 3 speed setting in User.vbd for JWC.  Then exchanged RAITH.JS and settimeout.js
scripts for scripts with extended pumpdown and Z motor time out limits. Tested load
unload three times with no problems.
Returned to Users access at 15:25:00