Subject: Problem raith SNF 2006-05-12 09:41:47: Z motor failure...
From: jwc@snf.stanford.edu
Date: Fri, 12 May 2006 09:41:48 -0700

Initial test of LOAD LOCK after recovery resulted in Z motor error in the ESCVIMTRL
module control.
Changes MOTOR3SPEED to 0.04 mm/sec in module ESCVIMTRL
Changes RAITH150.VBD MOTOR3SPEED TO 0.04 AND IN USER.VBD.
CHANGED SAMPLE STATUS TO REFLECT PROPER SAMPLE LOADER IN SYSTEM STATE (SAMPLESTAT 1)
FURTHER LL PROCEDURE TESTING IN PROGRESS.
SETTING UP FOR REPAIRS BY RAITH FIELD SERVICE WHEN PARTS BECOME AVAILABLE.  (EST. 5-25-06)
JAMES CONWAY
JWC