Subject: Comment raith SNF 2006-07-28 11:18:11: 30 kV test completed Users can use any ACCELERATION Voltage 0.3 to 30 KV
From: jwc@snf.stanford.edu
Date: Fri, 28 Jul 2006 11:18:13 -0700

Good Morning,
Yesterday we tested for the second time an entire afternoon of write operations at 30
kV with no problems.  Gun pressure base level is 5.2 E -010 Torr and system presuure
base is 1.9 E - 006 Torr.  Very stable gun emission and nominal beam currents levels
have been observed.  
Users can now utilize any ACCELERATION Voltage from 0.3 to 30 KV.
ALL SAMPLES GOING INTO THE SYSTEM MUST BE SOFT BAKED JUST BEFORE LOADING INTO RAITH.
Thank you for your support,