Subject: Problem raith SNF 2006-08-26 11:00:34: Beam current measurement
From: rohank@snf.stanford.edu
Date: Sat, 26 Aug 2006 11:00:35 -0700

The "measure" button on the beam current measurement was not 'active' (i.e., it was
greyed out). REstarting the Raoth-150 program didnt help. Upon talking to James, I followed
these steps:
1) Log out of Raith150 and Leo32 programs and close both programs
2) shutdown and restart Keithley 486 Picoammeter
3) Restart the  Leo and Raith and programs
Upon followong these steps, the problem was solved.