Subject: Comment raith SNF 2006-10-19 18:25:29: system is up and functional NOT TO EXCEED 10 KV
From: jwc@snf.stanford.edu
Date: Thu, 19 Oct 2006 18:25:30 -0700

Greetings:
All systems and modules are functional.
Users are not to go above 10 keV acceleration voltage as we had an EHT TRIP during the
first test at 15 kV.
Vacuum remains good at 5.8E-010T for gun and 1.6E-006 for chamber.
Good Night,