Subject: Problem raith SNF 2007-07-16 16:54:31: Problems with reduced water flow through Electron Optics board.
From: jwc@snf.stanford.edu
Date: Mon, 16 Jul 2007 16:54:32 -0700

Water flow error occurred at about 16:30 hours today. 
Users had been reporting starnge image jumping and drifting side to side over the weekend.
Checked chiller unit in basement chase and determined that the water flow loop had been
compromised and the water flow has been reduced to 0.4 liters per minute down from 0.8
to 1.2  lpm obesrved several weeks before.  Cooler outflow temperature is regulated
and in specification at 20 degrees C. -- return line at 23 - 24.8 degrees C. RAITH Field
service was called and left voicemail at 16:45 hours.  We have a rebuilt chiller in
house ready to install -- but likely will wish to replace turbo pump and e/o card components
and water supply and return lines at time of installation to prevent reoccurrance of
problem.
JWC