Subject: Problem raith SNF 2007-07-16 17:44:09: no 2 E/O card failure due to poor water flow
From: jwc@snf.stanford.edu
Date: Mon, 16 Jul 2007 17:44:09 -0700

17:15 RAITH SEM imaging and Electrostatic lens control funtions were lost and the system
went into standby and a partial vent of system chamber occurred.  Manually stopped the
vent by unplugginbg airlock cable and reset the entire hardware on the raith using the
yellow light power reset method.
Successfully brought up both computers -- started all hardware and firmware iincluding
restarted system chamber turbo and reestablished E/o card functionality.  tested SEM
imaging and handed off to the current User.
Users are requested to call JWC upon encountering any further problems this evening.
 RAITH has been contacted by phone.
JWC