Subject: Re: Shutdown raith SNF 2008-05-22 14:23:37: majors systems malfunction
From: jwc@snf.stanford.edu
Date: Thu, 22 May 2008 16:36:10 -0700

water flow error brought system down and it vented chamber to 500 torr and placed in
standby.
water flow at reduced levels at 0.4 - 0.8 lpm.  possibl;e blockage or obstruction in
cooling loop which was just rebuilt last friday.
Tested stage and Lasers , motors for current, and recalibrated stage positions and travel
limits.
system back up for users. Ready to go.