Subject: Re: Problem raith SNF 2008-05-27 10:06:35: Oops .... accidentally cleared problem ...
From: shott@snf.stanford.edu
Date: Fri, 30 May 2008 22:38:28 -0700

This was a problem related to the cooling-water induced vacuum problems and has been
resolved by finding and fixing a kink in the water line to the electronics.  Particularly
when the door closes, the line tended to kink.
John