Subject: Shutdown raith SNF 2008-06-10 10:18:30: turbo pump failed
From: jwc@snf.stanford.edu
Date: Tue, 10 Jun 2008 10:18:31 -0700

Lev markov with raith will be in shortly to evaluate.
I could not maintain turbo pumping to system.  likely a time to reach suitable speed
interlock failed.
system will be off line to users until replacement has been switched out in system.