Problem drytek2: Process Cooling Water Flow is Low
(
(none)
)
02:
Shutdown drytek2: drytek2 cooling water flow 0.2
(
(none)
)
03:
Shutdown drytek2:
(
(none)
)
04:
Problem drytek2:
(
(none)
)
05:
Problem drytek2: drytek2 is not venting
(
(none)
)
06:
Shutdown drytek2: Shutdown - chamber cannot vent to ATM
(
(none)
)
07:
Shutdown drytek2: cleared Shutdown - repaired chamber vent
(
(none)
)
08:
Problem drytek2:
(
(none)
)
09:
Problem drytek2: "O2 Plasma" comes up PINK rather than dark blue. Most likely either N2 is flowing in place of O2 or chamber is not purging properly.
(
(none)
)
10:
Problem drytek2:
(
(none)
)
11:
Problem drytek2:
(
(none)
)
12:
Shutdown drytek2: Lots of dust in chamber
(
(none)
)
13:
Shutdown Clear drytek2:
(
(none)
)
14:
Shutdown drytek2:
(
(none)
)
15:
Shutdown Clear drytek2:
(
(none)
)
16:
Problem Clear drytek2:
(
(none)
)
17:
Problem Clear drytek2:
(
(none)
)
18:
Shutdown drytek2: Len
(
booth.snf.stanford.edu
)
19:
Problem drytek2: Len 2
(
booth.snf.stanford.edu
)
20:
Problem Clear drytek2:
(
booth.snf.stanford.edu
)
21:
Shutdown Clear drytek2:
(
booth.snf.stanford.edu
)
22:
Shutdown drytek2: working on chiller¨
(
cbaxter.snf.stanford.edu
)
23:
Shutdown Clear drytek2:
(
cbaxter.snf.stanford.edu
)
24:
Problem drytek2: rf power too low,
(
franklin.snf.stanford.edu
)
25:
Shutdown drytek2: low rf output
(
cbaxter.snf.stanford.edu
)
26:
Problem Clear drytek2: part is installed and tested
(
cbaxter.snf.stanford.edu
)
27:
Shutdown Clear drytek2: part installed and tested
(
cbaxter.snf.stanford.edu
)
28:
Problem drytek2: No oxygen flow
(
usriniva.snf.stanford.edu
)
29:
Shutdown drytek2: no o2 flow...
(
cbaxter.snf.stanford.edu
)
30:
Shutdown Clear drytek2: mks controller was,it was bad contact on connector.
(
cbaxter.snf.stanford.edu
)
31:
Problem Clear drytek2: reseated connector on display
(
cbaxter.snf.stanford.edu
)
32:
Problem drytek2:
(
grupp.snf.stanford.edu
)
33:
Problem Clear drytek2: cleared problem - replaced 3-4 flow readout.
(
booth.snf.stanford.edu
)
34:
Problem drytek2: top electrode, back right, plastic pin is gone
(
sminne.snf.stanford.edu
)
35:
Problem drytek2: no rf power
(
sylviajs.snf.stanford.edu
)
36:
Shutdown drytek2: 500pa generator has a blown 7amp fuse. looking for a replacement...
(
cbaxter.snf.stanford.edu
)
37:
Shutdown Clear drytek2: changed shorted bridge rectifier and 7amp fuse
(
cbaxter.snf.stanford.edu
)
38:
Problem Clear drytek2: change shorted rectifier
(
cbaxter.snf.stanford.edu
)
39:
Problem Clear drytek2: installed missing teflon guide..
(
cbaxter.snf.stanford.edu
)
40:
Problem drytek2: flowmeter
(
michaud.snf.stanford.edu
)
41:
Problem Clear drytek2:
(
booth.snf.stanford.edu
)
42:
Problem drytek2: standby bulb is burned out
(
beckwith.snf.stanford.edu
)
43:
Problem Clear drytek2: replaced bulb.
(
booth.snf.stanford.edu
)
44:
Problem drytek2: Fixing pin missing on tray 3
(
chcheng.snf.stanford.edu
)
45:
Problem Clear drytek2: installed teflon tip..
(
cbaxter.snf.stanford.edu
)
46:
Problem drytek2: cooling system not operating
(
utkan.snf.stanford.edu
)
47:
Shutdown drytek2: no electrode cooling.
(
booth.snf.stanford.edu
)
48:
Shutdown Clear drytek2: changed fuse holder and fuse
(
cbaxter.snf.stanford.edu
)
49:
Problem Clear drytek2: change fuse
(
cbaxter.snf.stanford.edu
)
50:
Problem drytek2: Gives error E10 when trying to start process. Could not get standby to run! Steve
(
sminne.snf.stanford.edu
)
51:
Shutdown drytek2: E-10 error
(
booth.snf.stanford.edu
)
52:
Shutdown Clear drytek2: door was not closed.
(
booth.snf.stanford.edu
)
53:
Problem Clear drytek2: door not closed.
(
booth.snf.stanford.edu
)
54:
Problem drytek2: System light burned out
(
greene.snf.stanford.edu
)
55:
Problem Clear drytek2: changed light bulb
(
cbaxter.snf.stanford.edu
)
56:
Problem Clear drytek2: changed light bulb
(
cbaxter.snf.stanford.edu
)
57:
Problem Clear drytek2:
(
cbaxter.snf.stanford.edu
)
58:
Problem drytek2: RF on button burnt out.
(
latta.snf.stanford.edu
)
59:
Problem Clear drytek2: changed rf on indicator bulb
(
cbaxter.snf.stanford.edu
)
60:
Problem drytek2: power
(
mcp.snf.stanford.edu
)
61:
Shutdown drytek2: Electrode cooling water at zero flow
(
sminne.snf.stanford.edu
)
62:
Shutdown Clear drytek2: removed scale from electrodes.
(
booth.snf.stanford.edu
)
63:
Shutdown drytek2: Chilled H2O to electrodes is at zero.
(
latta.snf.stanford.edu
)
64:
Shutdown Clear drytek2: fixed chiller & cleared electrodes.
(
booth.snf.stanford.edu
)
65:
Shutdown drytek2: Electrode temperature 30degC.
(
booth.snf.stanford.edu
)
66:
Shutdown Clear drytek2: set chiller temp.
(
booth.snf.stanford.edu
)
67:
Shutdown drytek2: chiller repair.
(
booth.snf.stanford.edu
)
68:
Shutdown Clear drytek2: Changed to temporary electrode chiller.
(
booth.snf.stanford.edu
)
69:
Shutdown drytek2: need to changed dimister filter
(
cbaxter.snf.stanford.edu
)
70:
Shutdown Clear drytek2: demister changed completed
(
cbaxter.snf.stanford.edu
)
71:
Shutdown drytek2: drytek 2 cooling water flow = 0.
(
emes.snf.stanford.edu
)
72:
Shutdown Clear drytek2: restarted chiller.
(
booth.snf.stanford.edu
)
73:
Shutdown drytek2: Cooling electrode too low
(
pzwahlen.snf.stanford.edu
)
74:
Shutdown Clear drytek2: restarted water flow
(
booth.snf.stanford.edu
)
75:
Problem drytek2: O2 MFC errorDate: Nov 17, 2000 7:28:52 PM
(
pzwahlen.snf.stanford.edu
)
76:
Problem drytek2: O2 MFC error
(
booth.snf.stanford.edu
)
77:
Problem drytek2: pink powder near lowest electrode
(
amol.snf.stanford.edu
)
78:
Problem drytek2: pink powder near lowest electrode
(
cbaxter.snf.stanford.edu
)
79:
Shutdown drytek2: Vane pump needs oil.
(
booth.snf.stanford.edu
)
80:
Shutdown drytek2: Vane pump needs oil.
(
booth.snf.stanford.edu
)
81:
Shutdown drytek2: RF Power Does not turn on
(
utkan.snf.stanford.edu
)
82:
Shutdown drytek2: RF Power Does not turn on
(
latta.snf.stanford.edu
)
83:
Problem drytek2: Cooling electrode is low
(
pzwahlen.snf.stanford.edu
)
84:
Problem drytek2: Cooling electrode is low
(
booth.snf.stanford.edu
)
85:
Shutdown drytek2: chiller chang
(
cbaxter.snf.stanford.edu
)
86:
Shutdown drytek2: chiller chang
(
cbaxter.snf.stanford.edu
)
87:
Shutdown drytek2 2001-01-02 18:24:41: need to change oil
(
cbaxter.snf.stanford.edu
)
88:
Re: Shutdown drytek2 2001-01-02 18:24:41: need to change oil
(
cbaxter.snf.stanford.edu
)
89:
Problem drytek2 2001-01-03 12:49:37: No forward power
(
lindaw.snf.stanford.edu
)
90:
Shutdown drytek2 2001-01-03 13:51:49: No RF. Electrodes need to be flushed...
(
latta.snf.stanford.edu
)
91:
Re: Problem drytek2 2001-01-03 12:49:37: No forward power
(
cbaxter.snf.stanford.edu
)
92:
Re: Shutdown drytek2 2001-01-03 13:51:49: No RF. Electrodes need to be flushed...
(
cbaxter.snf.stanford.edu
)
93:
Problem drytek2 2001-01-04 00:04:15:
(
davehong.snf.stanford.edu
)
94:
Shutdown drytek2 2001-01-04 08:48:57:
(
booth.snf.stanford.edu
)
95:
Re: Shutdown drytek2 2001-01-04 08:48:57: installed rectifier and all three cooling fan. rf tested
(
cbaxter.snf.stanford.edu
)
96:
Re: Problem drytek2 2001-01-04 00:04:15: changed rectifier and cooling fan
(
cbaxter.snf.stanford.edu
)
97:
Problem drytek2 2001-01-11 22:53:58: load complete light does not go off
(
mbadi.snf.stanford.edu
)
98:
Re: Problem drytek2 2001-01-11 22:53:58: load complete light does not go off
(
booth.snf.stanford.edu
)
99:
Shutdown drytek2 2001-01-22 07:49:55: electrode cooling water flow below minimum.
(
booth.snf.stanford.edu
)
[<<-]
[<-]
[->]
[->>]
|