Subject: H2 regulator Issue
From: Ted Berg <tberg@stanford.edu>
Date: Thu, 11 Oct 2012 06:38:55 -0700
Hello All,
A quick update on the H2 regulator issue. The H2 regulator has
developed creep i.e. it does not stay where it is set it moves up in
pressure if there is no flow. When the system is in inert idle, the
creep causes the H2 to exceed allowed pressure and the chamber goes
offline and cannot be brought back until pressure drops. A replacement
regulator assembly is being prepared but will not be ready until next
week. Stay tuned we may have a temporary work around. ted