Subject: Problem evbond: evbond top chuck temp erratic
From: dlaser@snf.stanford.edu
Date: Nov 27, 2000 8:18:49 PM

While running newstd.aba on the evbonder this afternoon, I noticed that the temperature
reading of the top chuck fluctuated around the target temp of 350C by as much as +/-
60C.  Frequency of fluctation, although erratic, was fairly high, perhaps suggesting
a sensor problem; also, temp reading fluctuated most rapidly while the voltage was on.