Subject: Problem evbond 2002-07-22 17:59:55: flag pull not reliable
From: zappe@snf.stanford.edu
Date: Jul 22, 2002 5:59:56 PM

The flags were not pulled out reliably during the process although flag-in position
was indicated when the wafer and holder were loaded and the flags were properly inserted
between silicon and pyrex wafer.
Testing of the pull-procedure suggests that the pins are probably bent down and are
more pushing against the holder rather than pulling the flags.