01:
Shutdown metalica: Metalica cryopump was found at high temp. Starting cryo regen...
(
(none)
)
02:
Comment metalica: Metalica cryo regen
(
(none)
)
03:
Shutdown Clear metalica: Cryo regen
(
(none)
)
04:
Problem metalica: metalica pressure gauge
(
yiching.snf.stanford.edu
)
05:
Problem Clear metalica: repaired gauge
(
mramsey.snf.stanford.edu
)
06:
Shutdown metalica: cryo pump at 60
(
phimmer.snf.stanford.edu
)
07:
Shutdown Clear metalica: regen. complete
(
mramsey.snf.stanford.edu
)
08:
Shutdown metalica: shut down for cryo regen
(
king.snf.stanford.edu
)
09:
Shutdown Clear metalica: EricP regen'ed the cryo
(
king.snf.stanford.edu
)
10:
Shutdown metalica: cryo temp high. Needs regen
(
narya.snf.stanford.edu
)
11:
Shutdown Clear metalica: regen complete
(
mramsey.snf.stanford.edu
)
12:
Comment metalica: LN2 bucket is cracked¨
(
king.snf.stanford.edu
)
13:
Comment metalica:
(
wingoh.snf.stanford.edu
)
14:
Comment metalica:
(
wingoh.snf.stanford.edu
)
15:
Shutdown metalica: Roughing problem
(
king.snf.stanford.edu
)
16:
Shutdown Clear metalica: changed s60 pump to d60 pump(fomblin)
(
cbaxter.snf.stanford.edu
)
17:
Problem metalica: rough pump
(
guo.snf.stanford.edu
)
18:
Comment metalica: RP ok- observations
(
ericp.snf.stanford.edu
)
19:
Problem metalica: Metallica : Silver paste is almost finished
(
narya.snf.stanford.edu
)
20:
Re: Problem metalica: Metallica : Silver paste is almost finished
(
Robin King
)
21:
Problem Clear metalica: rough pump clear
(
king.snf.stanford.edu
)
22:
Problem Clear metalica: silver paste found
(
king.snf.stanford.edu
)
23:
Shutdown metalica 2001-02-15 12:03:36: Cryo Regen
(
king.snf.stanford.edu
)
24:
Problem metalica 2001-02-16 01:12:55: Cryo Regen complete
(
ericp.snf.stanford.edu
)
25:
Re: Shutdown metalica 2001-02-15 12:03:36: Cryo Regen
(
jhaydon.snf.stanford.edu
)
26:
Re: Problem metalica 2001-02-16 01:12:55: Cryo Regen complete
(
jhaydon.snf.stanford.edu
)
27:
Shutdown metalica 2001-02-22 09:51:51: regen cryopump
(
king.snf.stanford.edu
)
28:
Re: Shutdown metalica 2001-02-22 09:51:51: regen cryopump
(
king.snf.stanford.edu
)
29:
Comment metalica 2001-04-02 14:28:14:
(
dton.snf.stanford.edu
)
30:
Problem metalica 2001-04-12 13:47:08: gate valve
(
szou.snf.stanford.edu
)
31:
Re: Problem metalica 2001-04-12 13:47:08: gate valve
(
jhaydon.snf.stanford.edu
)
32:
Problem metalica 2001-04-29 12:04:36: plasma not igniting
(
rmzapata.snf.stanford.edu
)
33:
Problem metalica 2001-04-29 12:05:19: meter(vacuum guage) not working
(
rmzapata.snf.stanford.edu
)
34:
Shutdown metalica 2001-04-29 12:05:46: plasma not igniting
(
rmzapata.snf.stanford.edu
)
35:
Comment metalica 2001-04-30 19:34:10: plasma ok- target loading problem
(
ericp.snf.stanford.edu
)
36:
Re: Shutdown metalica 2001-04-29 12:05:46: plasma not igniting
(
jhaydon.snf.stanford.edu
)
37:
Re: Problem metalica 2001-04-29 12:05:19: meter(vacuum guage) not working
(
jhaydon.snf.stanford.edu
)
38:
Re: Problem metalica 2001-04-29 12:04:36: plasma not igniting
(
jhaydon.snf.stanford.edu
)
39:
Shutdown metalica 2001-05-27 11:44:07: cryo needs regen
(
pruitt.snf.stanford.edu
)
40:
Problem metalica 2001-05-28 17:52:44: cryo ok
(
ericp.snf.stanford.edu
)
41:
Re: Shutdown metalica 2001-05-27 11:44:07: cryo needs regen
(
jhaydon.snf.stanford.edu
)
42:
Re: Problem metalica 2001-05-28 17:52:44: cryo ok
(
jhaydon.snf.stanford.edu
)
43:
Shutdown metalica 2001-06-19 10:30:49: cryopump regen
(
jhaydon.snf.stanford.edu
)
44:
Re: Shutdown metalica 2001-06-19 10:30:49: cryopump regen
(
cbaxter.snf.stanford.edu
)
45:
Comment metalica 2001-07-14 11:19:33: power supply cooling fan does not work
(
wingoh.snf.stanford.edu
)
46:
Problem metalica 2001-07-14 11:20:03: power supply cooling fan does not work
(
wingoh.snf.stanford.edu
)
47:
Re: Problem metalica 2001-07-14 11:20:03: power supply cooling fan does not work
(
jhaydon.snf.stanford.edu
)
48:
Problem metalica 2001-08-16 14:09:29: the resistance between the electrodes is low
(
txx.snf.stanford.edu
)
49:
Problem metalica 2001-08-17 00:35:45:
(
wangdw.snf.stanford.edu
)
50:
Re: Problem metalica 2001-08-17 00:35:45: part is found.
(
mahnaz.snf.stanford.edu
)
51:
Re: Problem metalica 2001-08-16 14:09:29: the resistance between the electrodes is low
(
mahnaz.snf.stanford.edu
)
52:
Problem metalica 2001-08-21 12:24:45:
(
txx.snf.stanford.edu
)
53:
Shutdown metalica 2001-08-21 12:44:04: shut down for cryo regen
(
jhaydon.snf.stanford.edu
)
54:
Re: Shutdown metalica 2001-08-21 12:44:04: shut down for cryo regen
(
jhaydon.snf.stanford.edu
)
55:
Re: Problem metalica 2001-08-21 12:24:45:
(
jhaydon.snf.stanford.edu
)
56:
Shutdown metalica 2001-08-25 07:43:30: facility
(
rcrane.snf.stanford.edu
)
57:
Re: Shutdown metalica 2001-08-25 07:43:30: facility
(
rcrane.snf.stanford.edu
)
58:
Comment metalica 2001-09-04 16:09:21: New Log Sheets
(
jperez.snf.stanford.edu
)
59:
Problem metalica 2001-09-20 13:18:50: output not stable
(
wangdw.snf.stanford.edu
)
60:
Problem metalica 2001-09-24 16:17:45: Metalica
(
jperez.snf.stanford.edu
)
61:
Re: Problem metalica 2001-09-24 16:17:45: Metalica
(
jhaydon.snf.stanford.edu
)
62:
Re: Problem metalica 2001-09-20 13:18:50: output not stable
(
jhaydon.snf.stanford.edu
)
63:
Comment metalica 2001-10-11 19:14:33: Base pressure larger than 1E-6 torr
(
kyang.snf.stanford.edu
)
64:
Problem metalica 2001-10-12 21:15:21: Thermal Paste
(
flannery.snf.stanford.edu
)
65:
Re: Problem metalica 2001-10-12 21:15:21: Thermal Paste
(
jperez.snf.stanford.edu
)
66:
Shutdown metalica 2001-10-17 14:30:39: Cryo regen needed
(
jperez.snf.stanford.edu
)
67:
Re: Shutdown metalica 2001-10-17 14:30:39: Cryo regen needed
(
cbaxter.snf.stanford.edu
)
68:
Shutdown metalica 2001-10-22 11:07:34: Can't get 10 ohms on targets
(
jperez.snf.stanford.edu
)
69:
Shutdown metalica 2001-10-22 11:11:45: Can't get >10 mega ohms
(
jperez.snf.stanford.edu
)
70:
Problem metalica 2001-10-22 22:25:22: Resistance ok
(
ericp.snf.stanford.edu
)
71:
Re: Shutdown metalica 2001-10-22 11:11:45: Can't get >10 mega ohms
(
jhaydon.snf.stanford.edu
)
72:
Re: Shutdown metalica 2001-10-22 11:07:34: Can't get 10 ohms on targets
(
jhaydon.snf.stanford.edu
)
73:
Re: Problem metalica 2001-10-22 22:25:22: Resistance ok
(
jhaydon.snf.stanford.edu
)
74:
Problem metalica 2001-10-28 19:22:46: side target shot (1 Ohm resisitance only)
(
nchen.snf.stanford.edu
)
75:
Shutdown metalica 2001-10-29 10:04:22: power supply
(
jperez.snf.stanford.edu
)
76:
Problem metalica 2001-10-29 11:50:20: PS ok, vacuum sw bad.
(
ericp.snf.stanford.edu
)
77:
Re: Shutdown metalica 2001-10-29 10:04:22: power supply
(
jhaydon.snf.stanford.edu
)
78:
Re: Problem metalica 2001-10-29 11:50:20: PS ok, vacuum sw bad.
(
jhaydon.snf.stanford.edu
)
79:
Re: Problem metalica 2001-10-28 19:22:46: side target shot (1 Ohm resisitance only)
(
jhaydon.snf.stanford.edu
)
80:
Problem metalica 2001-11-05 21:01:17: can not strike plasma
(
txx.snf.stanford.edu
)
81:
Re: Problem metalica 2001-11-05 21:01:17: can not strike plasma
(
jhaydon.snf.stanford.edu
)
82:
Problem metalica 2001-12-05 13:40:34: Cryopump needs Regen.
(
jperez.snf.stanford.edu
)
83:
Problem metalica 2001-12-07 01:18:19: Cryo regen complete
(
ericp.snf.stanford.edu
)
84:
Re: Problem metalica 2001-12-07 01:18:19: Cryo regen complete
(
jhaydon.snf.stanford.edu
)
85:
Re: Problem metalica 2001-12-05 13:40:34: Cryopump needs Regen.
(
jhaydon.snf.stanford.edu
)
86:
Shutdown metalica 2001-12-12 09:34:53: System won't vent
(
jperez.snf.stanford.edu
)
87:
Re: Shutdown metalica 2001-12-12 09:34:53: System won't vent
(
jhaydon.snf.stanford.edu
)
88:
Shutdown metalica 2001-12-20 18:20:24: Cryo out for rebuild
(
ericp.snf.stanford.edu
)
89:
Re: Shutdown metalica 2001-12-20 18:20:24: Cryo out for rebuild
(
jhaydon.snf.stanford.edu
)
90:
Re: Shutdown metalica 2002-02-22 10:29:21: Cryo Temperture too high to use, 23 k
(
jhaydon.snf.stanford.edu
)
91:
Shutdown metalica 2002-02-27 13:22:32: cyro temp. is too high (19K)
(
qwang.snf.stanford.edu
)
92:
Re: Shutdown metalica 2002-02-27 13:22:32: cyro temp. is too high (19K)
(
cbaxter.snf.stanford.edu
)
93:
Problem metalica 2002-03-11 14:29:09: Cryo temperature is too high (19)
(
jperez.snf.stanford.edu
)
94:
Re: Problem metalica 2002-03-11 14:29:09: Cryo temperature is too high (19)
(
jhaydon.snf.stanford.edu
)
95:
Shutdown metalica 2002-03-29 10:05:11: the temp. of cryopump is higher than 18K
(
qwang.snf.stanford.edu
)
96:
Re: Shutdown metalica 2002-03-29 10:05:11: the temp. of cryopump is higher than 18K
(
jhaydon.snf.stanford.edu
)
97:
Problem metalica 2002-03-29 15:12:32: cryo being regend
(
jhaydon.snf.stanford.edu
)
98:
Re: Problem metalica 2002-03-29 15:12:32: cryo being regend
(
jhaydon.snf.stanford.edu
)
99:
Problem metalica 2002-04-16 17:12:28: cryo temp over 18K
(