Shutdown p5000etch: Wafer did not drop down on blade from Chamber A. 4 wafers are still inside
(
(none)
)
02:
Shutdown Clear p5000etch: Wafer cleared.
(
(none)
)
03:
Shutdown p5000etch: Wafer did not drop on the arm from Chamber B
(
(none)
)
04:
Shutdown Clear p5000etch:
(
(none)
)
05:
Shutdown p5000etch: System repeatedly mishandling wafers
(
(none)
)
06:
Shutdown Clear p5000etch: system ready for conditioning
(
cbaxter.snf.stanford.edu
)
07:
Shutdown p5000etch: wafer stuck in chamber B
(
yiching.snf.stanford.edu
)
08:
Shutdown Clear p5000etch: Clear: removed wafer from Ch. B oxide.
(
booth.snf.stanford.edu
)
09:
Shutdown p5000etch: shutdown for chamabr C wet clean..
(
cbaxter.snf.stanford.edu
)
10:
Shutdown Clear p5000etch: cha. C wet clean completed. cha. will need conditioning before ranning any process.
(
cbaxter.snf.stanford.edu
)
11:
Shutdown p5000etch: wafer stuck in chamber c
(
true.snf.stanford.edu
)
12:
Shutdown Clear p5000etch: removed wafers fr cha. c
(
cbaxter.snf.stanford.edu
)
13:
Shutdown p5000etch: wafer stuck in ch B
(
true.snf.stanford.edu
)
14:
Shutdown Clear p5000etch: Shutdown cleared
(
shott.snf.stanford.edu
)
15:
Shutdown p5000etch:
(
manjot.snf.stanford.edu
)
16:
Shutdown Clear p5000etch: P5000 Cham D wafer out
(
mcvittie.snf.stanford.edu
)
17:
Shutdown Clear p5000etch: p5000 Cham D wafer out
(
mcvittie.snf.stanford.edu
)
18:
Problem p5000etch: Ch D Slit Door Flaps
(
true.snf.stanford.edu
)
19:
Shutdown p5000etch: wafer stuck in ch. b
(
true.snf.stanford.edu
)
20:
Shutdown Clear p5000etch: removed wafer from chamber
(
cbaxter.snf.stanford.edu
)
21:
Problem Clear p5000etch: changed actuator
(
cbaxter.snf.stanford.edu
)
22:
Shutdown p5000etch: p5000 wafer stuck in elevator
(
dinkar.snf.stanford.edu
)
23:
Shutdown Clear p5000etch: Removed Wafer from loadlock jim
(
mcvittie.snf.stanford.edu
)
24:
Shutdown p5000etch: need to put oil in pump.
(
booth.snf.stanford.edu
)
25:
Shutdown Clear p5000etch: recycled B pump oil.
(
booth.snf.stanford.edu
)
33:
Problem p5000etch: possible water problem
(
king.snf.stanford.edu
)
34:
Problem p5000etch: Gryphon Argon MFC
(
king.snf.stanford.edu
)
35:
Shutdown Clear p5000etch: changed contactor. heater is @setpoint
(
cbaxter.snf.stanford.edu
)
36:
Problem Clear p5000etch: changed contactor
(
cbaxter.snf.stanford.edu
)
37:
Problem Clear p5000etch: wafers removed
(
cbaxter.snf.stanford.edu
)
38:
Problem Clear p5000etch:
(
cbaxter.snf.stanford.edu
)
39:
Problem p5000etch:
(
manjot.snf.stanford.edu
)
40:
Problem Clear p5000etch:
(
cbaxter.snf.stanford.edu
)
41:
Problem p5000etch: ChamberD vane pump needs repair.
(
booth.snf.stanford.edu
)
42:
Shutdown p5000etch: load lock won't ventDate: Nov 17, 2000 6:54:18 PM
(
true.snf.stanford.edu
)
43:
Shutdown p5000etch: load lock won't vent
(
booth.snf.stanford.edu
)
44:
Shutdown p5000etch: could not vent the load chamber-wafers stuck
(
dhelqaq.snf.stanford.edu
)
45:
Shutdown p5000etch: could not vent the load chamber-wafers stuck
(
booth.snf.stanford.edu
)
46:
Problem p5000etch: ChamberD vane pump needs repair.
(
cbaxter.snf.stanford.edu
)
47:
Shutdown p5000etch 2001-01-02 06:50:05:
(
booth.snf.stanford.edu
)
48:
Re: Shutdown p5000etch 2001-01-02 06:50:05:
(
booth.snf.stanford.edu
)
49:
Shutdown p5000etch 2001-01-09 17:09:25: Wafer stuck in Ch B
(
latta.snf.stanford.edu
)
50:
Re: Shutdown p5000etch 2001-01-09 17:09:25: Wafer stuck in Ch B
(
cbaxter.snf.stanford.edu
)
51:
Problem p5000etch 2001-01-11 18:07:05: Wafer did not drop on blade from chamber C
(
manjot.snf.stanford.edu
)
52:
Shutdown p5000etch 2001-01-11 18:07:25:
(
manjot.snf.stanford.edu
)
53:
Re: Shutdown p5000etch 2001-01-11 18:07:25: removed wafer out of chamber C
(
cbaxter.snf.stanford.edu
)
54:
Re: Problem p5000etch 2001-01-11 18:07:05: Wafer did not drop on blade from chamber C
(
cbaxter.snf.stanford.edu
)
55:
Shutdown p5000etch 2001-01-18 17:26:51: heat exchangers shutting off system
(
booth.snf.stanford.edu
)
56:
Re: Shutdown p5000etch 2001-01-18 17:26:51: heat exchangers shutting off system
(
cbaxter.snf.stanford.edu
)
57:
Shutdown p5000etch 2001-01-31 10:11:53: flowmeter on HX-1 failed.
(
booth.snf.stanford.edu
)
58:
Re: Shutdown p5000etch 2001-01-31 10:11:53: flowmeter on HX-1 failed.
(
cbaxter.snf.stanford.edu
)
59:
Problem p5000etch 2001-02-07 16:26:19: Ch. D alarm... I just cleared it.
(
latta.snf.stanford.edu
)
60:
Re: Problem p5000etch 2001-02-07 16:26:19: Ch. D alarm... I just cleared it.
(
cbaxter.snf.stanford.edu
)
61:
Problem p5000etch 2001-02-09 15:51:44: Chamber C, no HBr available
(
booth.snf.stanford.edu
)
62:
Shutdown p5000etch 2001-02-13 07:35:09: HX-1 flow fault is turning off heater circuit.
(
booth.snf.stanford.edu
)
63:
Re: Shutdown p5000etch 2001-02-13 07:35:09: HX-1 flow fault is turning off heater circuit.
(
booth.snf.stanford.edu
)
64:
Problem p5000etch 2001-02-14 09:08:04: Chamber A, Chlorine nearly empty
(
booth.snf.stanford.edu
)
65:
Re: Problem p5000etch 2001-02-14 09:08:04: Chamber A, Chlorine nearly empty
(
booth.snf.stanford.edu
)
66:
Problem p5000etch 2001-02-16 17:43:15: SYSTEM WILL NOT UNLOAD WAFERS FROM ELEVATOR TO CASSETTE, JIM MCVITTE UNLOADED THEM.
(
manjot.snf.stanford.edu
)
67:
Re: Problem p5000etch 2001-02-16 17:43:15: SYSTEM WILL NOT UNLOAD WAFERS FROM ELEVATOR TO CASSETTE, JIM MCVITTE UNLOADED THEM.
(
booth.snf.stanford.edu
)
68:
Shutdown p5000etch 2001-02-21 03:21:03: WAFERS STUCK INSIDE CHAMBER A. WILL NOT UNLOAD AFTER PROCESS, ERROR " CANNOT VENT LOAD CHAMBER ".
(
manjot.snf.stanford.edu
)
69:
Re: Shutdown p5000etch 2001-02-21 03:21:03: WAFERS STUCK INSIDE CHAMBER A. WILL NOT UNLOAD AFTER PROCESS, ERROR " CANNOT VENT LOAD CHAMBER ".
(
cbaxter.snf.stanford.edu
)
70:
Problem p5000etch 2001-02-27 21:08:13: P5000 faults
(
metzler.snf.stanford.edu
)
71:
Re: Problem p5000etch 2001-02-27 21:08:13: P5000 faults
(
cbaxter.snf.stanford.edu
)
72:
Re: Problem p5000etch 2001-02-09 15:51:44: Chamber C, has HBr okay to run
(
seymour.snf.stanford.edu
)
73:
Shutdown p5000etch 2001-03-12 11:41:15: WAFER FELL INSIDE CHAMBER C
(
manjot.snf.stanford.edu
)
74:
Re: Shutdown p5000etch 2001-03-12 11:41:15: WAFER FELL INSIDE CHAMBER C
(
booth.snf.stanford.edu
)
75:
Comment p5000etch 2001-03-13 09:19:16: Ch B
(
metzler.snf.stanford.edu
)
76:
Shutdown p5000etch 2001-03-15 17:57:27: need to cycled fomblin oil on pumps..
(
cbaxter.snf.stanford.edu
)
77:
Re: Shutdown p5000etch 2001-03-15 17:57:27: need to cycled fomblin oil on pumps..
(
cbaxter.snf.stanford.edu
)
78:
Problem p5000etch 2001-03-18 19:25:17: Ch B P5000
(
metzler.snf.stanford.edu
)
79:
Re: Problem p5000etch 2001-03-18 19:25:17: Ch B P5000
(
cbaxter.snf.stanford.edu
)
80:
Shutdown p5000etch 2001-03-21 13:30:15: Wafer in CHC
(
song.snf.stanford.edu
)
81:
Re: Shutdown p5000etch 2001-03-21 13:30:15: Wafer in CHC
(
booth.snf.stanford.edu
)
82:
Shutdown p5000etch 2001-03-21 20:02:20:
(
song.snf.stanford.edu
)
83:
Re: Shutdown p5000etch 2001-03-21 20:02:20:
(
cbaxter.snf.stanford.edu
)
84:
Shutdown p5000etch 2001-04-01 20:21:26: robot stuck? It won't get wafers.
(
hopcroft.snf.stanford.edu
)
85:
Re: Shutdown p5000etch 2001-04-01 20:21:26: robot stuck? It won't get wafers.
(
cbaxter.snf.stanford.edu
)
86:
Problem p5000etch 2001-04-03 14:43:11: Chamber D will only heat to 180C.
(
mcvittie.snf.stanford.edu
)
87:
Problem p5000etch 2001-04-04 10:45:28: Shutdown Chamber A
(
booth.snf.stanford.edu
)
88:
Problem p5000etch 2001-04-13 12:08:02: ch b debris
(
metzler.snf.stanford.edu
)
89:
Re: Problem p5000etch 2001-04-13 12:08:02: ch b debris
(
booth.snf.stanford.edu
)
90:
Problem p5000etch 2001-04-18 07:48:18: Shutdown Chamber D
(
booth.snf.stanford.edu
)
91:
Problem p5000etch 2001-04-27 18:09:40: chamber c: MFC control
(
myliao.snf.stanford.edu
)
92:
Re: Problem p5000etch 2001-04-27 18:09:40: chamber c: MFC control
(
booth.snf.stanford.edu
)
93:
Shutdown p5000etch 2001-05-03 11:26:15: WAFER BROKE INSIDE CHAMBER C
(
manjot.snf.stanford.edu
)
94:
Re: Shutdown p5000etch 2001-05-03 11:26:15: WAFER BROKE INSIDE CHAMBER C
(
booth.snf.stanford.edu
)
95:
Problem p5000etch 2001-05-03 15:05:11: Ch C needs to be seasoned before regular use.
(
booth.snf.stanford.edu
)
96:
Problem p5000etch 2001-05-03 18:56:54: Chamber C He leak
(
greene.snf.stanford.edu
)
97:
Re: Problem p5000etch 2001-05-03 18:56:54: Chamber C He leak
(
cbaxter.snf.stanford.edu
)
98:
Re: Problem p5000etch 2001-05-03 15:05:11: Ch C needs to be seasoned before regular use.
(
cbaxter.snf.stanford.edu
)
99:
Shutdown p5000etch 2001-05-10 17:37:52: Wafer stuck inside Ch.A
(
narya.snf.stanford.edu
)
[<<-]
[<-]
[->]
[->>]
|