 Please note: the minimum ICP Forward pressure noted here is different from what is in the software for the tool.  This is at the suggestion Process Engineering at PlasmaTherm.  Using lower than the suggested minimum setpoint will cause dangerously high spiking reflected power.

Parameter
Set Point
Soft Tolerance

Hard Tolerance

Min
Max
Units
Process Time
10.0 0 +/-
0 +/- 0 1.0e6 s
Over Etch Time Percent
0
10 % 20 % 0 100 %
Pressure/Position Evacuation Condition
#pressure
#position
- - - - - - -
Pressure 10.0 5 +/- 10 +/- 0 100 mTorr
Throttle Valve Position Set Point
0 10 % 20 % 0 100 %
Process End Condition
#Time
#Endpoint
- - - - - - -
Recipe Restart Sequence Name
All Process Sequences Names
- - - - - - -
Recipe Abort Sequence Name All Process Sequences Names - - - - - - -
Cl2
0 2.0
+/- 5.0
+/- 0 114.2
sccm
BCl3
0 1.0
+/- 2.5 +/- 0 119.55
sccm
O2
0 2.0 +/- 5.0
+/- 0 96.6
sccm
SF6
0 2.0 +/- 5.0 +/- 0 103.8
sccm
Ar
0
2.0
 +/- 5.0 +/- 0  98.1  sccm
CF4 0 2.0
 +/- 5.0  +/- 0  101.8  sccm
CH4
0 2.0
+/- 5.0 +/- 0  135.7 sccm
N2 0 4.0
+/-  10.0 +/- 0  194.0 sccm
He Cooler Mode
#pressure
#flow
- - - - - - -
He Cooler Pressure Setpoint
0 10 % 20 % 0 10000 mTorr
He Cooler Flow Setpoint 0 10 % 20 % 0 10.0 sccm
Bias RF Forward Setpoint
0 10
% 20
% 0 600.0 W
IPC Forward Power Setpoint
0
10 % 20 % 200
1200.0
W
 Bias Match Control Mode  #auto
#man->auto
#manual
#hold
#default
 -  -  -  -  -  -  -
 Bias Match Load Position  0  10 %  20  %  0  100.0  %
 Bias Match Tune Position  0  10 %  20  %  0  100.0  %
IPC Match Control Mode
#auto
#man->auto
#manual
#hold
#default
- - - - - - -
IPC Match Load Position
0 10 % 20 % 0 100.0 %
IPC Match Tune Position 0 10 % 20 % 0 100.0 %
Temperature Electrode Setpoint
25 5 +/- 10 +/- 10.0 60.0 C
Temperature Lid Setpoint 25 25 +/- 50 +/- 20.0 180.0 C
                 
Temperature Spool Setpoint 25 25 +/- 50 +/- 20.0 180.0 C
Execute if later Step Restarted
False - - - - - - -
External Detector Endpoint Recipe Name
#none
#test.ewr
- - - - - - -
No Endpoint found alarm severity
#error
#warning
#none
-  - - - - - -
 Camera Mode Setpoint
 #none
#manual
 -  -  -  -  -  -  -
 Bias RF Power Error Timeout
 10000  0  %  0  %  0  60000  ms
 IPC RF Reflected Power Error Timeout
 10000  0  %  0  %  0  60000  ms

Document Actions